38
SYMBOLS & TECHNICAL SPECIFICATIONS
Gas Inputs:
280 kPa (40.6 psi) to 600 kPa (87 psi), via Oxygen DISS
and Air DISS 69 kPa (10 psi) to 275 kPa (40 psi), via Oxygen DISS (O
2
Conserve Mode) 0 kPa (0 psi) to 69 kPa (10 psi), via 1/8” Barb
Gas Outputs:
ISO 5356, 22mm
Gas Flow
:
5 to 60 LPM in volume modes and peak flows > 80 LPM in
pressure modes, CPAP and demand flows.
Electrical Rating:
•
Operating Voltage: 12 VDC, 5.0 Amp max current draw
•
Input Voltage AC: 110-240 VAC, 50/60 Hz <1 Amp max current
draw (55 watts)
•
Input Voltage DC: 11 – 15 VDC, 5.0 Amp max current draw
•
Fuses: 2 Amp slow acting, 250 VAC rating
WARNING:
Use only a 12 VDC HC1221W battery or equivalent for the
internal battery. Improper function or damage may occur with the
wrong battery.
WARNING:
Use only the specified fuse. Improper fuse rating may
result in damage to the ventilator or electrical shock to the user.
WARNING:
Grounding reliability can only be achieved when the
equipment is connected to a hospital grade receptacle.
WARNING:
The CCV is not MRI compatible.
Inspiratory Time:
0.5 to 2.0 seconds in increments of 0.25 seconds,
accuracy: ±10%
Breaths Per Minute:
Range dependent on ventilation settings, accuracy:
±10%
O
2
Blending Capability:
21% to 100% O
2
in 1% increments, accuracy:
±12%
CPAP/PEEP Range:
0 to 25 cmH
2
O, Inadvertent PEEP <2 cmH
2
O
PIP (setting):
15 to 55 cmH
2
O, accuracy: ±2 cmH
2
O or 10% (whichever is
greater
)