Ventilator Alarms
In alarm conditions, the ventilator will
sound an audible tone (>70 dB), illuminate
three red lights in a horizontal line, and a
single red light next to the label for the
specific alarm condition.
To mute the audible alarm for 110 seconds,
press the Alarm Silence/Reset button. The alarm lights will stay on. Any
new alarm conditions will also trigger the audible alarm.
The alarm lights stay on until the condition has been resolved
and
the
reset button has been pressed. This means a self-correcting alarm
condition will leave a visible record for the clinical provider. To clear the
illuminated alarm indicator, press the Alarm Silence/Reset button.
System Failure Alarm:
Activates if communication between the
processors fail. This indicates that the ventilator is not operating or may
not be operating per the intended settings. When the processor fails,
the alarm sounds and a light will turn on. This alarm may not be
silenced.
WARNING:
Any ventilator with a System Failure Alarm may pose a
danger to the patient and must be removed from service
immediately.
Critical Battery Alarm:
Activates when about 20 minutes of battery life
remain, clears 10 seconds after adequate power is restored. This alarm
may not be silenced. See
CCV BATTERIES
for more information about
the CCV battery system.
Low O
2
Source Alarm:
Activates if the oxygen source pressure drops
below 2.7 to 2.5 Bar (275 – 255 kPa, 40 – 37 psi). Clears 8 seconds after
pressure is restored. The CCV will switch to using the internal
compressor, but the displayed %O
2
will not change. This alarm does not
activate if the %O
2
setting is 21% (Air only).
Low Air Source Alarm:
Activates if the compressed medical air source
pressure drops below 2.7 to 2.5 Bar (275 – 255kPa, 40 – 37 psi). Clears 8
seconds after pressure is restored. This alarm does not activate if the
%O
2
setting is 100% (O
2
only). This alarm will not be engaged if using
room air (21% O
2
) via the internal compressor.