Low Flow Source Setup Procedure
To use a concentrator or flowmeter as the CCV’s oxygen source, the
following steps must be performed:
1)
Turn the CCV off.
2)
Ensure that the O
2
hose is
securely connected to the
DISS port on the CCV and the
outlet port on reservoir.
3)
Connect the low flow oxygen
source to the reservoir inlet
barb hose fitting.
4)
Turn the source on and check
for leaks.
5)
Lower the %O
2
setting to 21%
and set the other ventilation
parameters as desired.
6)
Ensure that Control Lock is
not engaged and press O
2
Conserve, noting that the
green light above the O
2
Conserve button turns on.
7)
Increase the %O
2
setting until the red alarm light next to the
message “%O
2
Setting cannot be achieved in O
2
Conserve Mode”
is illuminated.
8)
Alter the settings to lower oxygen usage until the red alarm light
goes off.