82
User Manual
Data Length
— The oscilloscope triggers when the SDA data length is equal to
the value set in the
Byte Length
softkey.
4. If you have set the oscilloscope to trigger on an EEPROM Data Read condition:
Press
the
Limit Range
softkey to set the oscilloscope to trigger when data is
=
(equal
to),
<
(less than), or
>
(greater than) the data value set in the
Data1
softkey.
The oscilloscope will trigger on the clock edge for the Ack bit after the trigger event is
found. This data byte does not need to occur directly after the control byte. The
oscilloscope will trigger on any data byte that meets the criteria defined by the
Byte
Length
and
Data1
softkeys during a current address read or a random read or a
sequential read cycle.
5. If you have set the oscilloscope to trigger on a 7- bit address read or write frame
condition or a 10- bit read or write frame condition:
a. Press the
Addr
softkey and turn the
Universal Knob
to select the 7- bit or 10- bit
device address.
You can select from an address range of 0x00 to 0x7F (7- bit) or 0x3FF (10-bit)
hexadecimal. When triggering on a read/write frame, the oscilloscope will trigger
after the start, address, read/write, acknowledge, and data events occur. If don't
care is selected (0xXX or 0xXXX) for the address, the address will be ignored.
The trigger will always occur on the 17th clock for 7- bit addressing or 26th clock
for 10- bit addressing.
b. Press the
Data1
or
Data2
softkey and turn the
Universal Knob
to select the 8- bit
data pattern on which to trigger. You can select a data value in the range of 0x00
to 0xFF (hexadecimal). The oscilloscope will trigger after the start, address,
read/write, acknowledge, and data events occur.
c.
If don't care (0xXX) is selected for data, the data will be ignored. The trigger will
always occur on the 17th clock for 7- bit addressing or 26th clock for 10- bit
addressing. If you have selected a three- byte trigger, press the
Data2
value
softkey and turn the
Universal Knob
to select the 8- bit data pattern on which to
trigger.
6. If you have set the oscilloscope to trigger on data length condition:
a. Press
Address
to set the SDA address length 7bit or 10 bit.
b. Press
Byte Length
softkey and turn the
Universal Knob
to select the byte
length.
c. The range of the
Byte Length
is 1 to 12.
All Rights Reserved - General Polytronic Systems Ltd., London
www.gpslimited.com
Содержание GPS-1000X Series
Страница 2: ...All Rights Reserved General Polytronic Systems Ltd London www gpslimited com ...
Страница 105: ...User Manual 89 Figure 44 SPI Trigger All Rights Reserved General Polytronic Systems Ltd London www gpslimited com ...
Страница 114: ...98 User Manual Figure 54 CAN Trigger All Rights Reserved General Polytronic Systems Ltd London www gpslimited com ...