160
User Manual
To Set Aux Output
Users can set the type of the signal output from the
[Trigger Out]
connector at the rear
panel.
1. Press the
Utility
button on the front panel to enter the UTILITY function menu.
2. Press the
IO Set
softkey to enter the I/O SET function menu.
3. Press the Aux Output softkey to select Trig Out or Pass/Fail. The default setup is
Trig
Out
.
Trig Out
: if this type is selected, the oscilloscope outputs a signal that can reflect
the current capture rate of the oscilloscope at each trigger. The maximal capture
rate is 110k wfs/s.
Pass/Fail
: if this type is selected, the oscilloscope will output a pulse signal when
failed waveforms are detected. This signal can be connected to other control
systems to conveniently view the test results.
All Rights Reserved - General Polytronic Systems Ltd., London
www.gpslimited.com
Содержание GPS-1000X Series
Страница 2: ...All Rights Reserved General Polytronic Systems Ltd London www gpslimited com ...
Страница 105: ...User Manual 89 Figure 44 SPI Trigger All Rights Reserved General Polytronic Systems Ltd London www gpslimited com ...
Страница 114: ...98 User Manual Figure 54 CAN Trigger All Rights Reserved General Polytronic Systems Ltd London www gpslimited com ...