162
User Manual
To Update Firmware and Configuration
The firmware and configuration can be updated directly via USB flash driver.
Do the following steps to update the firmware:
1. Insert the USB flash driver which with the firmware and the configure files inside into
the USB host interface on the front panel of the oscilloscope.
2. Press the
Utility
button on the front panel to enter the UTILITY function menu.
3. Press the
Next Page
softkey to go to the second page of the UTILITY function menu.
4. Press the
Next Page
softkey to go to the third page of the UTILITY function menu.
5. Press the
Update
softkey to enter the UPDATE function menu.
6. Press the
Firmware
softkey and it will pop-out the message “
Press ‘Single’ to
continue and press ‘Run/Stop’ to exit!
”
7. Press the
Single
button to enter the SAVE/RECALL file system.
8. Turn the
Universal Knob
to select the update file which should be with an ADS postfix;
and then press the
Load
softkey to start update the firmware. The process needs
about 7 minutes. And during the update, do not cut off the oscilloscope, otherwise the
oscilloscope will not restart again.
9. After finish the update, the screen will pop-out the message “
Update success, please
restart the DSO
”, and you will heard a sound of buzzer.
10. Restart the oscilloscope to finish the firmware update.
Do the following steps to update the configuration:
1. Insert the USB flash driver which with the firmware and the configure files inside into
the USB host interface on the front panel of the oscilloscope.
2. Press the
Utility
button on the front panel to enter the UTILITY function menu.
3. Press the
Next Page
softkey to go to the second page of the UTILITY function menu.
4. Press the
Next Page
softkey to go to the third page of the UTILITY function menu.
5. Press the
Update
softkey to enter the UPDATE function menu.
6. Press the
Configure
softkey to enter the SAVE/RECALL file system.
7. Turn the
Universal Knob
to select the update file which should be with an CFG
postfix; and then press the
Load
softkey to start update the firmware. The process
needs about 30 seconds.
8. After finish the update, the screen will pop-out the message “
Update success, please
restart the DSO
”, and you will heard a sound of buzzer.
9. Restart the oscilloscope to finish the configuration update.
All Rights Reserved - General Polytronic Systems Ltd., London
www.gpslimited.com
Содержание GPS-1000X Series
Страница 2: ...All Rights Reserved General Polytronic Systems Ltd London www gpslimited com ...
Страница 105: ...User Manual 89 Figure 44 SPI Trigger All Rights Reserved General Polytronic Systems Ltd London www gpslimited com ...
Страница 114: ...98 User Manual Figure 54 CAN Trigger All Rights Reserved General Polytronic Systems Ltd London www gpslimited com ...