26
User Manual
To Specify Probe Attenuation Factor
Set the probe attenuation factor to match the type of the probe that you are using to
ensure correct vertical readouts.
Press the
CH1
button on the front panel; then press the
Probe
softkey and turn the
Universal Knob
to select the desired value and push the knob to confirm. The default
setup is
1X
.
The current probe attenuation factor is displayed in the channel label at the right side of
the screen. You can also press the
Probe
softkey
continuously to switch the probe
attenuation factor.
The table shows the probe attenuation factor
Menu
Attenuation Factor
0.1X
0.1 : 1
0.2X
0.2 : 1
0.5X
0.5 : 1
1X
1 : 1
2X
2 : 1
…
…
5000X
5000 : 1
10000X
10000 : 1
To Specify channel Input Impedance
The channel input impedance matching gives you the most accurate measurements
because reflections are minimized along the signal path.
Impedance setting to 1MΩ is for use with many passive probes and for general-
purpose measurements. The higher impedance minimizes the loading effect of the
oscilloscope on the device under test.
Impedance setting to 50Ω matches 50 ohm cables commonly used in making high
frequency measurements, and 50 ohm active probes.
Press the CH1 button on the front panel; then press the
Impedance
softkey to select the
desired impedance.
The current channel input impedance is displayed in the channel label at the right side of
the screen.
All Rights Reserved - General Polytronic Systems Ltd., London
www.gpslimited.com
Содержание GPS-1000X Series
Страница 2: ...All Rights Reserved General Polytronic Systems Ltd London www gpslimited com ...
Страница 105: ...User Manual 89 Figure 44 SPI Trigger All Rights Reserved General Polytronic Systems Ltd London www gpslimited com ...
Страница 114: ...98 User Manual Figure 54 CAN Trigger All Rights Reserved General Polytronic Systems Ltd London www gpslimited com ...