GMC-I Messtechnik GmbH
53
9.7
Recalibration
The measuring tasks performed with your instrument, and the
stressing it’s subjected to, influence aging of its components and
may result in deviation from the specified levels of accuracy.
In the case of strict measuring accuracy requirements, as well as
in the event of use at construction sites with frequent stress due
to transport and considerable temperature fluctuation, we recom-
mend a relatively short calibration interval of once per year. If your
instrument is used primarily in the laboratory and indoors without
considerable climatic or mechanical stressing, a calibration inter-
val of once every 2 to 3 years is sufficient as a rule.
During recalibration* at an accredited calibration laboratory
(DIN EN ISO/IEC 17025), deviations from traceable standards
demonstrated by your measuring instrument are documented.
Ascertained deviations are used to correct display values during
later use of the instrument.
We would be happy to perform DAkkS or factory calibration for
you at our calibration laboratory. Further information is available
on our website:
www.gossenmetrawatt.com (
COMPANY
Quality and Certif-
icates
DAKKS Calibration Center
Calibration Questions and
Answers.
Recalibration of your instrument at regular intervals is essential for
the fulfillment of requirements according to quality management
systems per DIN EN ISO 9001.
Note
Regular calibration of the test instrument should be per-
formed in a calibration laboratory which is accredited per
DIN EN ISO/IEC 17025.
* Examination of the specification, as well as adjustment, are not included
in calibration. However, in the case of our own products, any required
adjustment is performed and adherence to the specification is con-
firmed.
9.8
Manufacturer’s Guarantee
All
METRA HIT
digital multimeters and calibration instruments are
guaranteed for a period of 3 years after date of shipment. The
manufacturer’s guarantee covers materials and workmanship.
Damages resulting from use for any other than the intended pur-
pose or operating errors, as well as any and all consequential
damages, are excluded.
The calibration certificate confirms that the product’s specified
technical data were complied with at the point in time of calibra-
tion. We guarantee compliance with the specified technical data
within the permissible tolerances for a period of 12 months after
delivery.
10
Accessories
10.1
General
The extensive accessories available for our measuring instru-
ments are checked for compliance with currently valid safety reg-
ulations at regular intervals, and are expanded as required for new
applications. Currently up-to-date accessories which are suitable
for your measuring instrument are listed at the following web
address along with photo, order number, description and,
depending upon the scope of the respective accessory, data
sheet and operating instructions: www.gossenmetrawatt.de (
PRODUCTS
Measuring Technology – Portable – Multimeters
Specialized Applications
METRAHIT IM XTRA
...
Accessories).
10.2
Technical Data for Measurement Cables
(included with KS17-2 cable set and Z270S probe)
Electrical Safety
Observe the measuring instrument’s maximum values for electrical
safety!
Ambient Conditions (EN 61010-031)
Temperature
–20
C ... + 50
C
Relative humidity
50 ... 80%
Pollution degree
2
Use of KS17-2 and Z270S
Attention!
!
Measurements per DIN EN 61010-031 may only be per-
formed in environments in accordance with measuring
category III with the safety cap attached to the test probe
at the end of the measurement cable.
In order to establish contact inside 4 mm jacks, the safety caps
have to be removed by prying open the snap fastener with a
pointed object (e.g. the other test probe).
Maximum rated voltage
600 V
1000 V 1000 V
Measuring category
CAT IV
CAT III
CAT II
Maximum rated current
1 A
1 A
16 A
With safety cap attached
—
Without safety cap
—
—