13.11 U – Probe Voltage
➭
Connect the DUT to the test socket.
➭
Contact the ungrounded output for
safety extra-low voltage with test probe
P1.
➭
Select line voltage polarity.
Special Case: Permanently Installed DUT
➭
Contact all voltage conducting parts with
test probe P1.
Meas. Parameter
Meaning
Measurement type
PE – P1
Measurement of voltages with reference
to PE, test socket remains voltage-free,
for permanently connected DUTs
PE – P1 (with mains)
Measurement of voltages with ref-
erence to PE, line voltage is ap-
plied to the test socket
Polarity
Normal
Reversed
Selection of polarity for mains volt-
age to the test socket
(only for PE-P1 with mains)
Select measuring function
1
U
Select parameters
2
Set parameters
3
Connect the DUT
4
PE – P1 (with mains)
Probe voltage
P1
PE – P1
Probe voltage
P1
Start test
5
Acknowledge line voltage warning
6
!
Only for meas. type:
(with mains)
Save measured values to clipboard
7
Stop test
8
Save measurements under ID no.
9