8.3
User Interface Icons – Parameter and Softkey Icons
Icon
Setup
Page
Parameters and Their Significance
Complete overviews of all icons are included in the full operating instructions.
1/3
All measurements:
Ref. voltage: Voltage to which measured leakage current values are
standardized; residual current protection: value for residual current monitoring (10/30 mA)
1/3
Auto measurements
: Set test sequence parameters: start and end view, consider
measuring uncertainty (yes/no), auto measuring point (yes/no)
1/3
Database
:
deletion,
statistics,
with inserted USB stick
:
Database:
backup,
restore
1/3
System
: set general device parameters:
date/time,
brightness,
volume,
default settings,
self-test
2/3
Printer
: printer selection for USB master port
connected,
disconnected
2/3
Inspector
:
select inspector from list,
add new inspector
2/3
Culture
: select language for user interface, keyboard and measuring sequences
by acknowledging the respective national flag; required!
2/3
Optionally connected external devices
USB stick,
keyboard / barcode scanner,
printer
3/3
System information
: query software and hardware version, serial number, build
number, calibration data and memory occupancy
—
Functions and their significance
Set classification parameters for the respective test sequence
(test sequences at switch positions A1 to A9)
Accept parameters, acknowledge message
Cancel single measurement or test sequence
Evaluate measurement of visual inspection with
OK
or
not OK
(toggle key)
Continue test, next test step in the test sequence
Left icon: Direct selection key for measuring type (connection type ...) or measuring
method (direct ...) / Right icon: Selection between two statuses (no submenu)
Start evaluation – record measured value. Each time this softkey is pressed, an
additional measured value is saved and the number is increased by one.
Left icon: Repeat measured value recording
Right icon: Repeat test step
Left icon: Delete measured value
Right icon: Skip individual tests in a test sequence
Display measured values from performed measurements and test sequences
Magnifying glass icon:
show (+) or hide (–) details regarding database objects or
selected measurements
Enter a new ID for a test object either before or after a test, and in case the ID
has not yet been entered to the structure
Save measurement data / save measurement data as (with display of directory
path / ID or new entry of an ID other than the preselected one)