Gossen Metrawatt GmbH
21
SECU
TEST ST
B
ASE
(
10
) /
P
RO
and SECU
LIFE ST
BASE
(25)
Example
Configure the standard designations in the
SECUTEST ST PRO.
Save the setting with the green checkmark.
The “OVE E 8701” designation is then used
in the test instrument.
9.2.2
Configuring Rotary Switch Positions
We recommend assigning frequently used
test sequences to A1 through A8 and
reserving switch position A9 for special
sequences, for which parameters often need
to be adjusted.
➭
Select an orange rotary switch position
(
A1
…
A9
), after which the start page for
the respective test sequence is displayed
(i.e. the integrated test sequence set at
the factory).
➭
Select classification parameters.
The
Classification Parameters 1/2
page appears.
The standard which is currently assigned
to the respective rotary switch position is
displayed under
Standard / Test Sequence
.
➭
Select
Standard / Test Sequence
.
The
Standard / Test Sequence
page ap-
pears. All standards available on the test
instrument are displayed here according
to the selected setting (off / national des-
ignation – see section 9.2.1).
➭
Select the desired standard.
The menu is returned to the
Classification
Parameters 1/2
page.
➭
Finally, confirm with the green
checkmark.
The settings are saved.
➭
Repeat the above described procedure
for each of the respective rotary switch
positions.
SETUP 1/3 > Auto Measurements 4/7 >VDE 0701-0702