9.15 Remote Sensing
219
Resolution
0.1 V
Voltage setting limit 2 (other than Voltage setting limit 1)
GKP-2303
Positive
voltage
Setting Range
(Peak value)
+0.1 V to +220.0 V / +0.1 V to +440.0 V
Factory default
+220.0 V / +440.0 V
Negative
voltage
Setting Range
(Peak value)
220.0 V to
10.0 V /
440.0 V to
20.0 V
Factory default
220.0 V /
440.0 V
Resolution
0.1 V
Note: The limitation is applied to the additional values of the AC voltage setting (recalculated
to a peak value) and the DC voltage setting.
Frequency setting limit (the lower limit
the upper limit)
GKP-2303
Upper
limit
Setting Range
1.0 Hz (AC mode : 40 Hz) to 550.0 Hz
Factory default
550.0 Hz
Lower
limit
Setting Range
1.0 Hz (AC mode : 40 Hz) to 550.0 Hz
Factory default
1.0 Hz (AC mode : 40 Hz)
Resolution
0.1 Hz
Note: In the AC mode, the setting range is 40.0 Hz to 550.0 Hz.
9.15
Remote Sensing
This switches the voltage used for measurement. When the remote sensing is on, the sensing
input terminal voltage is used. When it is off, the output terminal voltage is used.
By combining with AGC or Autocal, a voltage drop due to wiring to the load can be
compensated. When the remote sensing is on, the output voltage detection point corrected by
the AGC or Autocal function is switched to the sensing input terminal. When AGC or Autocal
is off, only the detection voltage used for measurement display is switched.
Effective only for AC-INT, AC-VCA, AC-SYNC, DC-INT, and DC-VCA and when the
waveform is sine wave or DC. It is turned off when the Sequence or Simulation is selected.
Measurement
voltage, power, power factor
AGC/Autocal
Off
On
On
Use the sensing input terminal
voltage
Not active
Active
Off
Use the output terminal voltage
Not active (factory
default)
Active
9.16
AGC
When the AGC (Automatic Gain Control) is on, the detection point voltage is always measured,
and the output voltage is continuously corrected so that its effective value is equal to the output
Содержание GKP-2302
Страница 15: ...1 1 OUTLINE 1 1 Overview 2 1 2 Features 2...
Страница 28: ...GKP 2302 User Manual 14 Nothing is connected to the output terminal...
Страница 60: ......
Страница 186: ......
Страница 187: ...173 5 DESCRIPTION OF SCREEN AND MENU 5 1 Screen Configuration 174 5 2 Menu Composition 177...
Страница 195: ...181 6 REMOTE CONTROL 6 1 Communication Interface 182 6 2 Remote Local State Switching 188...
Страница 203: ...189 7 TROUBLESHOOTING 7 1 Error Messages and Error Handling 190 7 2 When a Failure is Suspected 196...
Страница 216: ......
Страница 222: ......
Страница 245: ...9 34 Outline Dimensional Drawing 231 9 34 Outline Dimensional Drawing Figure 9 2 Outline Dimensional Drawing...