GKP-2302 User Manual
146
2.
In the [Auto Cal] item, select [1: OFF]. When [OFF] is selected here, the Autocal calibration
finishes, and the calibration factor is cleared.
3.
Use the [Rmt Sense] item to set the Remote sensing function to on or off, if necessary.
4.
Close the window.
4.14
Adjusting DC Offset
Even when the output voltage is set to 0 V, the DC offset voltage from several mV to several
tens mV may exist in the output. The DC offset adjustment function can reduce such DC
voltage close to zero.
For the DC offset adjustment, four types of values are retained for each output voltage range
and AC/DC mode. The setting range of the DC offset adjustment value is shown in Table 4-11.
Table 4-11 Setting Range of DC Offset Adjustment Value
AC/DC mode
Minimum
value
Maximum
value
Resolution
Initial value
Unit
AC
50.0
+50.0
0.1
0.0
mV
ACDC, DC
250
+250
1
0
mV
Note: The setting range is common to 100 V range and 200 V range.
Содержание GKP-2302
Страница 15: ...1 1 OUTLINE 1 1 Overview 2 1 2 Features 2...
Страница 28: ...GKP 2302 User Manual 14 Nothing is connected to the output terminal...
Страница 60: ......
Страница 186: ......
Страница 187: ...173 5 DESCRIPTION OF SCREEN AND MENU 5 1 Screen Configuration 174 5 2 Menu Composition 177...
Страница 195: ...181 6 REMOTE CONTROL 6 1 Communication Interface 182 6 2 Remote Local State Switching 188...
Страница 203: ...189 7 TROUBLESHOOTING 7 1 Error Messages and Error Handling 190 7 2 When a Failure is Suspected 196...
Страница 216: ......
Страница 222: ......
Страница 245: ...9 34 Outline Dimensional Drawing 231 9 34 Outline Dimensional Drawing Figure 9 2 Outline Dimensional Drawing...