10 Servicing of the OSMOMAT auto
■
Adjusting the Thermistor probe
The thermistor probe must be checked for centricity in the lower cooling system:
1. After work on the upper and/or lower measurement system,
2. In the event of erroneous and/or deviating measurement results,
3. During the quality assurance routine.
At this point only a description of menu functions is provided. The actual measurement system testing
and setting procedure will not be available until the next version of this user guide.
Select >
adjust
>
thermistor probe
in the initial display. Confirm with
OK
to initiate start-up or shutdown
of the thermistor probe.
Start-up and shutdown are controlled by a stepper motor. The length of the period for which the button
is pressed determines the number of steps.
1. Select
up
to start up the thermistor probe incrementally.
2. Select
down
to shut down the thermistor probe incrementally.
3. Press
Esc
to return the thermistor probe to its preset position
Please note: The automatic lift limit of the elevator is invalidated during this operation.
This can result in overwinding of the excenter disk and destroy the measurement
system!
38 | Servicing of the OSMOMAT auto