10 Servicing of the OSMOMAT auto
This function can be switched off if not needed.
Select >
adjust
>
cleaning paper feed
in the initial display. Confirm with
OK
.
According to the current setting, you now have the option of activating the function by selecting
on
or
deactivating it by selecting
off
.
■
Checking the Initiation Function
The initiation function must be checked following replacement of a defective initiation needle, for
example.
To check the initiation function, select >
adjust
>
actuate cryst motor
in the initial display. Confirm with
OK
to rotate the initiation needle one time.
■
New Initialization
Reinitialize the OSMOMAT auto if additional calibrations are unsuccessful after an unsuccessful or
cancelled calibration.
The memory content and the settings for the date and time are retained after re-initialization.
Select >
adjust
>
new initialization
in the initial display. Confirm with
OK
to conduct a new initialization.
The system restarts automatically.
Select
cancel
to cancel the operation.
Servicing of the OSMOMAT auto
| 37