8 Memory
Management
8 Memory
Management
The OSMOMAT auto has an internal battery-buffered CMOS memory for 300 pieces of measurement
data. The data is retained in the memory even after the unit is switched off or disconnected from the
power supply. The manufacturer-specified battery life is 10 years. After expiration of this time, the
battery buffer may not be sufficient after the unit is switched off or disconnected from the power supply
to retain the data. The data is then permanently lost.
■
Memory Space Assignment
To obtain an overview of the instantaneous memory space assignment, select >
sample
>
memory
>
memory capacity
in the initial display. Confirm your selection with
OK
.
The amount of available space
is indicated in positions. (One position per sample measurement).
Select
next
to return to the display >
memory capacity
. Use
Esc
to return to the display >
memory
and press
Esc
again to return to the initial display.
■
Searching for Batch and Sample Results
You can search for sample results in the memory by batch. Select >
sample
>
memory
>
look at
charge
in the initial display. Confirm your selection with
OK
. The last batch is displayed or
"there is no
data in memory"
is displayed when there is no data in the memory. Use
Esc
to return to the display >
memory
.
The batch data and the sample volume are displayed for every batch.
Press the
button to display the batches in decreasing chronological order.
3
-
Press the
button to display the batches in increasing chronological order.
2
+
Confirm your selection with
OK
to display the sample results for the batch.
You can end this operation at any time via the
Esc
key.
The date, sample position in the sample holder, and the measurement result in osmol/kg are shown
for every sample measurement.
Press the
button to display the samples in increasing chronological order.
2
+
Press the
button to display the samples in decreasing chronological order.
3
-
You can end this operation at any time via the
Esc
key.
Memory Management
| 25