Page 16 of 51
OSMOMAT 050 User Guide
Version 1.1 (2011-10-12)
The tubes are pushed onto the stainless steel tubes by about
5 mm
. The tool for widening the tube
ends (50.9. 0140) included with the standard accessories is used for this purpose. The tube end is
held using sandpaper and widened using the moistened tip of the tool.
The connection tubes for the supply and waste bottles are passed through a bore of the unscrewed
cap and pushed onto the corresponding plungers
(Figure 4-2).
The plungers hold the tubes on the
bottom of the bottles. The second opening on the cap provides unrestricted atmospheric pressure
compensation.
Figure 4-2: Plunger
Tube lengths
Connection tube between supply bottle and pump
59 cm
Connection tube between ground bolt and waste bottle
45 cm
Connection tube between pump and measuring cell
6 cm
Connection tube between measuring cell and ground bolt
5 cm
Pump tube
9,5 cm
Table 4-1 Tube lengths