OSMOMAT 050 User Guide
Version 1.1 (2011-10-12)
Next, immediately remove the ethanol using a paper tissue
(Figure 4-7)
and add distilled water.
If the alcohol remains in the measuring cell too long, it affects the plastic material, which be-
comes opaque.
.
Use the tuberculin syringe again to aspirate the distilled water into the bore in the lower half of the
measuring cell. Watch closely to ensure that no air bubbles form. Next, remove the water from the
measuring cell using a paper tissue.
Figure 4-8: Adding the degassed solvent
The lower half of the measuring cell is now cleaned, and you can add the solvent.
Use the 30 mL syringe to add the degassed Ringer's solution
(Figure 4-5)
.
After adding the degassed
solvent, rinse the bore of the lower half of the measuring cell using the tuberculin syringe again. Dur-
ing injection, do not push the syringe plunger all the way to the end to prevent air bubbles from being
pushed into the bore.
Figure 4-6: Filling the lower measuring cell
Figure 4-7: Cleaning the lower measuring cell