4. Unscrew and remove lower measuring cell by loosening the two screws below the housing.
5. Remove the lower measuring cell and the large O ring (63.22 x 1.78 mm).
6. Loosen the counter nut and unscrew and remove the pressure transducer from the measuring
cell.Remove the small O ring from the measuring cell and do not reuse under any
circumstanes
7. Thoroughly rinse the lower half of the measuring cell using water and dry.