4 Preparing the Measuring cell for Measurement
To prepare the measuring cell, you need the tools and accessories included with the standard acces-
sories as well as a suitable solvent, distilled water and the prepared semipermeable membrane and
filter paper
(see chapter 4.4)
.
4.1 Making the Tube Connections during Initial Startup
The tubes included with the standard accessories are connected according to
Figure 4-1: Making the
tube connections
.
Figure 4-1: Making the tube connections
Use only the Tygon tubes supplied by the manufacturer. Other tubing material is per-
meable to air, which causes air bubbles that may severely impact or prevent the
measurement.
Bore for pressure
compensation
Ringer's solution
Supply bottle
plunger
Waste, minimum
fill level 1 cm
Waste bottle
plunger
Connection tube between
ground and waste bottle
Connection tube between
supply bottle and pump
Ground
Connection tube between measuring cell
and ground
Upper measuring cell
T screw
Lower measuring cell
Septum screw
Tube nipples
Connection tube between pump
and measuring cell
Tube pump
Pump tube