8.1 Flushcontrol Parameters
Programs Menu
Setting of Flushtime, Interval and Relaxtime
The
Flushtime
is the duration of flushing, i.e. the time for which relays K1 and K3 close. In the flushtime no
new measurement will be acted upon.
The
Interval
is the time between the beginning of a flush and the beginning of the next flush. If flu
latencytime is greater than the interval, then the next flush will occur 1 min after the relaxtime.
The
Relaxtime
is the time after the end of the flushtime in which the sensor can adjust to the surrounding
medium.
In this time no new measurement will be acted upon.
Aborts the setting and switches to the Pro-
grams Menu.
Aborts the input and switches to
the Programs Menu.
Deletes the last entered character.
Saves the settings and switches
to the Programs Menu.
Saves the input and switches to the next
menu. Without setting the old value is stored.
The settings of the Flushcontrol are saved and completed.
Presently active values are shown as “old value“.