15
Replacing the fuses
With a Phillips screwdriver, unscrew the lid (09) of the electronic board compartment, then remove
the broken fuse and place the new fuse (19) in the position indicated.
WARNING:
Before installing or replacing fuses, unplug the electric power cord.
uses that come with the Autoclave:
- 2 15A fuses installed in the equipment (F2/F3) + 1 spare;
- 1 08A fuse installed in the equipment (F4) + 1 spare;
- 1 0.5A fuse installed in the equipment (F1) + 1 spare.
WARNING:
- After performing the first replacement, it is advisable to always have spare fuses at hand.
- All fuses must be installed independently from the grid voltage.
Preparing the autoclave
Connect and fix the rubber hose (10) using the clamp
(11) to the external steam outlet (12) found in the
autoclave’s rear part. The other end must be placed in
a water container (20) or in a drainage which can stand
high temperatures.
In order to prevent the obstruction of water and steam
flow, keep the hose straight, avoiding the formation of
bends and siphons.
WARNING
Use only the hose supplied by Gnatus, or one
similar, so as to prevent if from melting when in
contact with steam, which may cause the external
steam outlet obstruction.
INSTALLATION OF EQUIPMENT