2) Submit the autoclave sterilization process in the TNT tissue used as primary packaging material
for sterile products and / or drapes. TNT is the acronym for (non-woven fabric), that is, a fabric
produced from disoriented fibers that are bonded and fixed, not going through the most common
textile processes that are spinning and weaving.
This tissue is not manufactures to be used in autoclaves, as TNT has polypropylene as the main raw
material, which is similar to a thermoplastic.
Warning
Gnatus is not responsible for any equipment damage or sterilization process due to the use
of kraft and/or TNT paper.
Observe each country’s standards related to wrapping preparation.
Note
Gnatus recommend the use of surgical paper, because the microbial barrier of this material
is above 90%, and it permits the operator to visually check the state of the instruments.
It also has chemical indicators of the process.
Gnatus shall not be responsible for:
- Use of the equipment differing from that for which it is intended.
- Damages caused to the equipment, the professional and/or the patient by the incorrect installation
and erroneous procedures of maintenance, differing from those described in these Instructions for
use which come with the equipment or by the incorrect operation of it.
Precautions and warnings “after” the use of equipment
- Turn off the main switch of the dental set when it is not in use for an extended period of time.
- Always maintain the equipment clean for the next operation.
- Do not modify any part of the equipment. Do not disconnect the cable or other connections
without need.
Precautions and warnings during the “cleaning and disinfection” of equipment
- Before cleaning the equipment, turn off the main switch.
- Avoid spilling water, even accidentally, or other liquids inside the equipment, which could cause
short circuits.
- Do not use microabrasive material or steel wool when cleaning, or employ organic solvents or
detergents which contain solvents such as ether, stain remover, etc.
Precautions in case of alteration in the functioning of equipment
- If the equipment has any abnormality, check if the problem is related to any item listed in the
topic of unforeseen events (failures, causes and solutions). If it is not possible to resolve the
problem, turn off the equipment, remove the power supply cable from the socket and contact your
representative (Gnatus).
Precautions to be adopted against foreseeable or uncommon risks, related to the
deactivation and abandoning of equipment
In order to avoid environmental contamination or undue use of the Equipment after it has become
useless, it should be discarded in the suitable place (as per the local legislation of the country).
Pay attention to the local legislation of the country for the conditions of installation and disposal
of residue.
PRECAUTIONS, RESTRICTIONS AND WARNINGS