Operating manual for the TG basic 1 appliance tester
Gilgen, Müller & Weigert
Measuring the voltage at an external earthed mains socket
Measuring range:
50…270 V AC
Display range:
50...270 V AC
Resolution:
1 V
Tolerance:
±±±±±
5 %
General technical data:
Power supply:
6 x 1.5 V batteries
Typ IEC LR6 (AA)
6 x 1.2 V NiMH
rechargeable batteries
Degree of pollution:
2
Overvoltage category:
CAT II 300 V
Protection degree:
IP40
Protection class:
II
Electrical safety:
EN61010-1/VDE0411
EMC emission:
EN61000-6-3
EMC interference resistance:
EN61326-1
Dimensions (L x W x H):
Approx. 265 x 110 x 50 mm
Weight:
Approx. 700 g
12. Notes on cleaning
When dirty, the tester is to be cleaned with a dry cloth (no
solvents) or an anti-static cloth.
The tester must be protected from shock and impacts.
13. Calibrating the tester
According to the DIN VDE 0701-0702:2008-06 standards:
„Measuring devices used for testing must be regularly
checked and calibrated.“
We recommend a calibration interval of one year. Shorter periods
are recommended if the tester is frequently used or used under
tough conditions. If the tester is seldom used then the calibration
interval can be extended up to a maximum of 3 years.
28
Note
: The device was developed, built and inspected in
accordance with DIN VDE 0404 Parts 1, 2 and DIN VDE 0413,
Parts 1, 2, 4