Operating manual for the TG basic 1 appliance tester
Gilgen, Müller & Weigert
After selecting the validity by briefly pressing the „Start“ button
(< 2 s), the desired selection is confirmed by pressing the „Start“
button for a longer time (> 2 s), after which you then reach the
following menu:
Here, you select the length of the protective conductor of the
test object („Start“ button < 2 s). The limit value resulting from this
selection is displayed in the upper status bar. The selected limit
value is then confirmed by pressing the „Start“ button for a
longer time (> 2 s).
The maximum limit value that can be set is
≤≤≤≤≤
1 Ohm. The limit
value is not permanently stored. This means that the limit value is
reset to
≤≤≤≤≤
0.3 Ohm when the device is manually switched off via
the rotary switch or when the device automatically switches off
after approx. 3 minutes.
Valid for protection
Valid for cable test
Valid for protection
class I and
cable test