Operating manual for the TG basic 1 appliance tester
Gilgen, Müller & Weigert
9.1.4. Test assessment
When the automatic test sequence finishes, an assessment
window is displayed.
The measurements and their assessments („X“ or „
3
3
3
3
3
“) are
displayed here.
The status bar displays „Test PASS“ or „Test FAIL“ depending on
the result.
Briefly pressing the „START“ button will repeat the test or perform
a new test if a new test object has been connected. Pressing the
„START“ button for longer than 2 s will display the quick guide
(see point 6.2.)
9.1.3. Measuring the substitute leakage current
The substitute leakage current is measured automatically.
Switching to permanent measurement is not possible.
Exceeding of the limit value is indicated as follows:
- The momentary measurement value blinks
- An audio signal is emitted
21