7. Specifications
GME LinScan Revision 14, dated 2016-06-08
Page 46 of 46
No.
Message
Cause
User Action
Circuit
ponent defective
vice again. If error per-
sists contact service.
527
Thermoelement Sensor 2: Short
Circuit
Hardware or software com-
ponent defective
Switch off and on de-
vice again. If error per-
sists contact service.
528
Laser pulse short error occurred.
Please restart treatment!
Hardware or software com-
ponent defective
If error only occurs
once then continue
working. If error occurs
frequently contact ser-
vice.
534
Synchronization timeout error oc-
curred
Hardware or software compo-
nent defective
Switch off and on device
again. If error persists
contact service.
598
Applicator System Tick Error
Hardware or software com-
ponent defective
Switch off and on de-
vice again. If error per-
sists contact service.
599
Applicator detects an Program
CRC error
Hardware or software com-
ponent defective
Switch off and on de-
vice again. If error per-
sists contact service.
7.5 Accessories
The following accessories/ detachable parts are intended to be used with the medical product:
x
Foot switch (Item-No. GME1020.2201)
x
Protective goggles (Item-No. GME1042.2100)
x
Applicator holder (Item-No. GME1040.2102)
x
Small spot adapter (Item-No. GME1040.4000)
x
Cover Door Interlock (Item-No. GME1020.0124)
x
Power cable (Item-No. GME1040.2200)
x
Transport box LinScan (Item-No. GME1040.3000)
x
Refill kit (Item-No. GME1040.2002)