7. Specifications
GME LinScan Revision 14, dated 2016-06-08
Page 36 of 46
Fragile, handle with care
Consult Instructions for Use
Do not stack
Avoid moisture
Made in Germany
7.3 Electromagnetic compatibility (EMC)
The medical device was tested according to EN 60601-1-2 on electromagnetic compatibility. Never-
theless, it should be noted that RF communication devices can affect electrical medical devices.
Therefore, the medical device must be installed and operated strictly according to the prescribed envi-
ronmental and installation instructions. RF devices must be operated at a safe distance, which de-
pends on the frequency and the power of the RF device.
In addition, the following precautions should be observed:
Caution!
Hazard:
The EMC tests were performed with the device and the
cables and connectors specified by the manufacturer and included
in the delivery of your device. Using other cables or connectors
can negatively affect the EMC of the device. When using such ca-
bles or connectors other devices placed in the surrounding could
be disturbed or the other devices could disturb the medical device.
Action
: Only use the original cable and / or connectors supplied
by the manufacturer.
Caution!
Hazard:
When testing the device for compliance with EN 60601-1-
2 standardized measuring distances are used. These distances
may be greater than the distances to nearby electrical devices that
occur when devices are placed directly next to, above or below the
medical device. With a too tight set-up it cannot be excluded that
the electrical devices interfere with each other and thus a proper
functioning with respect to the intended use is no longer guaran-
teed. This does not apply to equipment that the manufacturer has
specified and tested for such a placement.
Action:
Do not place electrical devices in the immediate vicinity of
the medical device, except for accessories that were released for
this purpose by the manufacturer.
!
!