5. Device Operation
GME LinScan Revision 14, dated 2016-06-08
Page 28 of 46
Figure 7: Schematic drawing
–
fixing the small spot adapter
Remove the adapter from the sapphire crystal after you finished treatment and deactivate the adapter
mode in the scan menu by choosing another scan area size.
Caution!
Hazard:
If the adaptor is not attached and the software is in the
adaptor mode, a higher fluency will be delivered.
Action:
Attach adaptor then select adaptor mode in software. De-
select adaptor mode when adaptor is removed.
Caution!
Hazard:
There will be no laser emission if the adapter is fixed in
the wrong direction to the sapphire crystal.
Action:
Please check whether the screw of the adapter points in
the direction of the applicator handle before using the small spot
adapter.
5.8 Menu Individual Parameters: Storage of Treatment Parameters
To save the last set of your treatment parameters, press the "Save" button on the treatment screen.
Pressing this button takes you to the "Individual Parameters" menu. Here you can save the data using
a name of your own choice or load already stored parameters.
To enter the menu “Individual Parameters” directly press the button “Stored settings” in main menu.
By pressing the button of a previously stored parameter set you enter the treatment menu. The stored
parameters are now already set. By confirming with the "Laser Start" button you can directly start the
treatment with this set of parameters.
!
!