Error
message
Causes
Measures
Remarks
0 error
Error cannot be identified.
Switch off the device and
restart.
If this error message appears
repeatedly, have the device
checked by the manufacturer.
1 error
Internal function test failed.
Switch off the device and
restart.
If this error message appears
repeatedly, have the device
checked by the manufacturer.
2 error
Segmentation fault
Switch off the device and
restart.
If this error message appears
repeatedly, have the device
checked by the manufacturer.
3 error
Information on the code chip
could not be read.
Remove the code chip and
insert again.
If this error message should
keep appearing, try to code the
device with another code chip
or have the device checked by
the manufacturer.
Error
message
Causes
Measures
Remarks
4 error
Coding error in the device.
Remove the code chip and
insert again.
If this error message should
keep appearing, try to code the
device with another code chip
or have the device checked by
the manufacturer.
5 error
Test strip isn‘t correctly
positioned or was removed
during measurement.
Remove the test strip from
the device and reinsert it.
Then repeat the measure-
ment.
6 error
Measurement is carried out
with a used test strip.
Remove the used test strip
from the device and repeat
the measurement with a new
test strip.
7 error
Bad operating conditions:
The ambient temperature is
too high.
Adapt the ambient tempera-
ture to the required operating
conditions (see technical
data).
Er
ro
r m
es
sa
ges
/c
au
ses
o
f e
rr
or
a
nd
m
eas
ur
es
to
ta
ke
6