background image

18

19

Step 2: 

In order to select the other unit of measure-

ment, press the MEM button. 

Step 3: 

If you are finished making settings, you can 

exit the settings menu by pressing the On/Off 

button. Otherwise, continue making settings 

by pressing the SET button.

2.6  Configuring the measured value 

memory

Note: 

The blood glucose meter stores the measured 

blood glucose values. The measured value 

memory can be configured so that the 

measurements can be allocated to eight 

different users. 

To do this, the measured value memory must 

be enabled for the individual users.

2.6.1  Configuration of the model variant 

BGM205

In its delivered condition, the measured  

value memory is only enabled for one user 

(“User 1”). 

The configuration of the measured value 

memory can be found after the unit of 

measurement setting in the settings menu 

sequence. 

Step 1:

To enable (or block) several measured value 

memories, press the SET button until the 

word “user” appears on the display in the up-

per right-hand corner (see figure), and to the 

left of the word “user”, the number “1” ap-

pears. In the lower area of the display, either 

the word “on” or “off” is blinking in addition. 

This indicates whether the measured value 

memory for user 1 is enabled or blocked at 

this time. 

mem

gluco-comfort

biocomfort

user

mem

gluco-comfort

biocomfort

mg/dL

mmol/L

G

et

tin

g s

ta

rt

ed

2

Содержание BGM 105

Страница 1: ...gluco comfort Blood glucose measuring systems BGM105 BGM 205 Operating manual Issued in 11 2006 REF 590107...

Страница 2: ...the blood glucose measurement 28 3 4 Carrying out the blood glucose measurement 31 3 5 Use by multiple persons 35 4 Measured value memory 36 4 1 Storing the measured values 36 4 2 Calling the stored m...

Страница 3: ...attery blinking dead battery 5 Number eld for Year Time User activated not activated Wireless connection activated not activated 6 Symbol for active wireless connection BGM 105 7 Code no indication wh...

Страница 4: ...is operating manual contains important information for you It is mandatory that the safety measures mentioned in this operating manual be taken The Biocomfort blood glucose measure ment system may onl...

Страница 5: ...ons for the test strips The test strips could be damaged by effects of temperature and light For this reason they should be used immediately after being removed from the test strip con tainer The test...

Страница 6: ...tion regarding the control solution Please observe the separate user instructions for the control solution Always shake the control solution well before use The control solution can be damaged if expo...

Страница 7: ...e Step 6 Afterwards check whether the battery has been correctlyinsertedbybrie ypressingtheOn Off button Inthiscase allelementsofthedisplayap pear for about two seconds as a function check Note Please...

Страница 8: ...lue you desire Now you can go to the next parameter by pressing the SET button or you can exit the settings menu by pressing the On Off button Note If you keep the MEM button pressed you can change th...

Страница 9: ...acti vate the minute display Step 10 Keep pressing the MEM button until the cor rect minute is displayed Step 11 If you are nished making settings you can exit the settings menu by pressing the On Of...

Страница 10: ...1 Con guration of the model variant BGM205 In its delivered condition the measured value memory is only enabled for one user User 1 The con guration of the measured value memory can be found after th...

Страница 11: ...ou wish you can set the status for user 2 by pressing the SET button Remark If for example you want to change the meas ured value memory status for user 5 press the SET button until the number 5 appea...

Страница 12: ...Off button Otherwise continue making settings by pressing the SET button 2 8 Clearing the measured value memory The operation for clearing the measured value memory is the last item in the settings m...

Страница 13: ...re for your blood glucose measurement as follows Step 1 Wash your hands thoroughly with warm water and soap Step 1 With the meter switched off insert the code chip included in the test strip package i...

Страница 14: ...ically Step 5 The user identity number of the measured value memory where the result of the last blood glucose measurement was stored blinks on the display see Section 3 5 Use by multiple persons Step...

Страница 15: ...ng the On Off button without inserting a test strip in the test strip receptacle Steps 3 6 are car ried out the same way as described before As opposed to the rst way with the inserted test strip the...

Страница 16: ...lease button Step 9 Carefully press on the nger pad until a blood drop is formed 3 4 Carrying out the blood glucose measurement Step 1 Hold the drop of blood next to the right side of the test strip A...

Страница 17: ...eeps The measured blood glucose value is displayed with the date and time of the measurement This data is stored in the measured value memory of the user Attention Hyperglycaemia warning blood glucose...

Страница 18: ...t strip out of the meter after the measurement Properly dispose of used test strips and lan cets immediately after measurement so that they cannot injure infect anyone 3 5 Use by multiple persons When...

Страница 19: ...the result of the last blood glucose measurement was stored see Section 3 5 Use by multiple persons Step 3 In order to look at the contents of the measured value memory of a certain user press the ME...

Страница 20: ...le measurement Step 8 To exit memory mode switch the device off again with the On Off button Once the end of the measured value memory has been reached or if there are no other measured values availab...

Страница 21: ...uring ranges can be checked with the different control solutions Normal blood glucose levels are simulated with the gluco comfort control solution GCS105 Low blood glucose levels are simulated with th...

Страница 22: ...for protecting and storing the supplies 5 7 Disposing of the meter When making blood glucose measurements blood can get onto and into the meter There fore a danger of infection is posed by used meters...

Страница 23: ...he device with another code chip or have the device checked by the manufacturer Error message Causes Measures Remarks 4 error Coding error in the device Remove the code chip and insert again If this e...

Страница 24: ...nt After restarting the device insert the test strip into the device Error message Causes Measures Remarks 16 error Blood glucose level too high out of measuring range of gluco comfort If the measured...

Страница 25: ...Housing is grey aluminium buttons and battery compartment cover are black BGM205 Housing is slate grey buttons and battery compartment are light blue Automatic switch off function 4 minutes when test...

Страница 26: ...GCS205 tar get value about 40 mg dL gluco comfort high control solution GCS305 target value about 400 mg dL Transportation and storage conditions 4 C to 30 C at a relative humidity of 10 to 95 When u...

Страница 27: ...tion are in vitro diagnostic supplies for personal use The CE marking of this product con rms that the relevant require ments of the EC directive 98 79 EC for these products are ful lled The blood glu...

Страница 28: ...54 55...

Отзывы: