Note: If you are not able to solve an occurring problem,
contact Biocomfort Diagnostics GmbH & Co. KG.
7 Technical data
Device type:
Blood glucose meter
Model:
gluco-comfort BGM105/
gluco-comfort BGM205
Serial no.:
Specified on the type
plate (bottom of measur-
ing device)
Measuring
method:
Glucose is oxidised by
a glucose dehydrogenase,
amperometric biosensor
Sample:
fresh capillary blood
Sample volume:
< 2 µl
Measuring range: 20 – 600 mg/dl
(1.1 – 33.3 mmol/l)
Measuring period: 10 seconds
Coding:
Code chip
Measured value
memory:
Max. 122 blood glucose
measurements per user,
all values are stored with
the time and date.
Average values:
7-, 14-, 21-, and 28-day
average
User management:
1 to 8 users
Size:
70 x 90 x 18
(W x L x H) [mm]
Weight:
circa 70 g
(without battery)
Colour:
BGM105: Housing
is grey aluminium,
buttons and battery
compartment cover are
black
BGM205: Housing is
slate grey, buttons and
battery compartment
are light blue
Automatic switch-off
function:
4 minutes when test
strip is inserted;
2 minutes if no func-
tion has been activated
and no test strip is
inserted.
Te
ch
ni
ca
l d
at
a
7
Error
message
Causes
Measures
Remarks
23 error
(BGM 105)
The wireless connection
could not be cut.
Switch off the device and
restart.
Measurement cannot be carried
out.
24 error
(BGM 105)
Communication error
Only occurs with the software,
which is available as an acces-
sory.
32 error
Function test of the system
was not completed success-
fully.
Switch off the device and
restart.
If this error message appears
repeatedly, have the device
checked by the manufacturer.
error and
blinking bat-
tery symbol
Battery is almost completely
used up.
Exchange the battery and
restart the device.
When the battery symbol blinks,
the blood glucose meter cannot
be used.