UM Ovens and Incubators EN rev. 0 31.05.2021
PARAMETER
DESCRIPTION
Pb
By modifying this parameter, it is possible to correct the reading of the PT100 temperature sensor
inside the instrument to a single temperature point. The correction will therefore be referable to
only one specific point.
PK
By modifying this parameter, it is possible to correct the reading of the PT100 temperature sensor
inside the instrument over the entire temperature range, i.e., it is possible to vary the inclination
of the reading range of the sensor itself.
PA
By modifying this parameter, it is possible to correct the reading of the PT100 room temperature
sensor installed on the instrument (refrigerated versions only) to a single temperature point. The
correction will therefore be referable to only one specific point.
14.
Clean and maintenance
A correct maintenance and cleaning of the instrument ensure that it remains in good condition.
The internal chamber of the instrument is made of stainless steel, so it can be cleaned with any detergent provided,
that is not aggressive and/or corrosive.
DANGER
Risk of electric shock and death
DO NOT spray water or cleaning agents onto the interior and exterior surfaces.
Switch off the main switch and pull out the mains plug before cleaning.
➢
Dry the instrument completely before switching it on again.
It is recommended that internal and external surfaces are cleaned with a normal all-purpose cleaner sprayed onto
a soft dampened cloth, so that it is not used in concentrated form. Before proceeding with cleaning or
decontamination, the user must ensure that the method used does not damage the instrument.
ATTENTION
Danger of corrosion
–
Damage to the instrument
DO NOT use cleaning agents containing acids or halides.
DO NOT use neutral cleaning agents on other surfaces
(e.g., on galvanised parts of the hinges or the rear wall of the housing).
Eye contact
–
Eye damage caused by chemical burns
DO NOT discharge into the sewage system.
➢
Wear safety goggles.
IMPORTANT:
If the instrument is to be sent for service, it should be properly cleaned and possibly decontaminated from
pathogens. It is also advisable to return the instrument in its original packaging to the repair service and if this is not
possible to pack it adequately for transport. Any damage caused by incorrect shipment will not be covered by
warranty.