Giorgio Bormac ICF-120 Plus Скачать руководство пользователя страница 11

 
 

10 

 

UM Ovens and Incubators EN rev. 0 31.05.2021 

 

Transportation of an already used oven/incubator 

 

 

Switch off the Argolab oven/incubator by pressing the main switch. 

 

Remove the power plug from the socket.

 

 

Remove the shelves.

 

 

Clean the Argolab oven/incubator and its shelves (see chapter 13 on page 18).

 

 

Dry the inside of the Argolab oven/incubator and the shelves. 

 

 

Wrap the shelves with bubble wrap. 

 

 

Pack the shelves in their original packaging and place them in the Argolab oven/incubator. 

 

Pack the Argolab oven/incubator in its original packaging. 

 

Take care that the Argolab oven/incubator does not get wet during transport. 

 

During transport, maintain the permitted room temperature (from -10 °C to 60 

°C). 

 

6.

 

Conservation

 

 

Store the Argolab oven/incubator only in closed, dry rooms. 

 

The  permitted  storage  temperature  is  -10  °C  to  60  °C.  The  maximum  permitted  storage  humidity  is  85%  RH 
without condensation. 

 

7.

 

First installation

 

 

Getting started

 

 The oven should be installed in following conditions

 

Dry, clean, stable worktable with a flat horizontal surface and heat resistant.  

 

At least 30 cm free around the instrument. 

 

Room temperature between 5 °C and 40 °C and relative humidity maximum of 85%. 

 

Power supply socket with earth connection. 

 

Power supply of 220/240 V - 50 Hz. 

 

                                    ATTENTION

 

  

Danger of overturning of the upper instrument 

 Damages to the instrument  

 

When stacking two instruments, place rubber pads under the four feet of the upper 
instrument.  

 

                                                                                                             

DANGER

 

 

Risk of explosion and death 

 

DO NOT operate the device in potentially explosive areas. 

 

DO NOT use explosive dust or air-soluble mixtures in the environment.

 

 

                                   ADVICE

 

 

Risk of overheating - Damage to the instrument 

   

 

DO NOT install the instrument in not ventilated area.

 

Ensure that there is enough ventilation to disperse heat.

 

Содержание ICF-120 Plus

Страница 1: ...OVENS and INCUBATORS User manual...

Страница 2: ...troller Model Description Temperature range ICN 16 Plus Natural convection incubator 16L maximum volume Room temperature from 5 C to 70 C ICN 35 Plus Natural convection incubator 35L maximum volume Ro...

Страница 3: ...gns 5 Symbols of obligation 5 2 GENERAL SAFETY INSTRUCTIONS 6 3 CE MARKING DATA 8 4 CONTENT OF PACKAGE 9 5 TRANSPORTATION 9 Instructions for safe transportation 9 Transportation of an already used ove...

Страница 4: ...Cycle functions 15 Repetition of a selected program 15 Delay starts function 16 12 INTRODUCTION OF SAMPLES INTO OVEN INCUBATOR 16 Limit temperature for sample protection 16 Restart mode after absence...

Страница 5: ...ions or incorrect results due to failure to comply with them Read entirely and carefully this manual and be sure to familiarize with the tool before starting to work with it This manual must be kept n...

Страница 6: ...surfaces Danger of damage to health caused by toxic substances Danger of injury from overturning of objects Risk of injury from lifting heavy objects Danger of environmental damage Danger of corrosio...

Страница 7: ...y where there is no risk of explosion Do not keep air solvent mixtures or explosive dusts nearby Never introduce materials into the instrument that are explosive or flammable at the selected operating...

Страница 8: ...f each other Always fix 2 stacked ovens with the fixing plates provided Risk of injury Risk of slipping or overturning and risk of damage to the instrument The instrument must be lifted by 2 people Th...

Страница 9: ...ors EN rev 0 31 05 2021 3 CE marking data Argolab instruments are manufactured in compliance with Directive 2006 42 EC and the relevant Community Directives applicable at the time of placing on the ma...

Страница 10: ...rtation ATTENTION Risk of injury of slipping or overturning and risk of damage to the instrument The instrument must be lifted by 2 people The instrument must be transported in its original packaging...

Страница 11: ...rmitted storage temperature is 10 C to 60 C The maximum permitted storage humidity is 85 RH without condensation 7 First installation Getting started The oven should be installed in following conditio...

Страница 12: ...11 UM Ovens and Incubators EN rev 0 31 05 2021 8 Instrument parts Display and commands Display Adjustment keys ON OFF Key Opening closing handle...

Страница 13: ...mperature variation at 150 C 0 3 C 0 3 C 0 4 C 0 5 C Heating time at 150 C 20 min 24 min 30 min 50 min Timer 99 59 hh min e 99 59 hh min e 99 59 hh min e 99 59 hh min e Safety class 3 1 3 1 3 1 3 1 Po...

Страница 14: ...at 37 C 40 min 45 min 55 min Timer 99 59 hh min e 99 59 hh min e 99 59 hh min e Safety class 3 1 3 1 3 1 Power supply power 230 V 600 W 230 V 700 W 230 V 1500 W Internal dimensions L A P 520 x 530 x 4...

Страница 15: ...using the buttons in H High M Medium L Low otherwise go to STEP 2 STEP 2 Use the and SHIFT keys to set the temperature value of the second working STEP Briefly press the SET PROG button to confirm the...

Страница 16: ...e this operation has been carried out the word Lk lock appears in the top right hand corner of the display in place of the word TIME next to the digits 0000 password Here below you can find the passwo...

Страница 17: ...mit for sample protection The instrument foresees the possibility of limiting the maximum working temperature to protect the samples from an incorrect temperature setting of the heating cycle Follow t...

Страница 18: ...ormal and therefore reducing the AL value excessively would risk causing the instrument to go into alarm frequently and unnecessarily Follow the instructions in paragraph 10 and use the keys to set th...

Страница 19: ...corrosive DANGER Risk of electric shock and death DO NOT spray water or cleaning agents onto the interior and exterior surfaces Switch off the main switch and pull out the mains plug before cleaning D...

Страница 20: ...does not apply to any product or parts thereof that have been damaged due to incorrect installation improper connection misuse accident or abnormal operating conditions No liability is accepted for d...

Отзывы: