![Giorgio Bormac ICF-120 Plus Скачать руководство пользователя страница 13](http://html1.mh-extra.com/html/giorgio-bormac/icf-120-plus/icf-120-plus_user-manual_2228777013.webp)
12
UM Ovens and Incubators EN rev. 0 31.05.2021
9.
Technical specifications
Natural convection ovens
TCN-30 Plus
TCN-50 Plus
TCN-115 Plus
TCN-200 Plus
Volume
30 l
50 l
115 l
200 l
Max temperature/ Resolution
+200/0,1°C
+300/0,1°C
+300/0,1°C
+300/0,1°C
Homogeneity at 150°C
± 3,5°C
± 3,5°C
± 3,5°C
± 4,0 °C
Temperature variation at 150°C
± 0,5°C
± 0,5°C
± 0,5°C
± 0,7°C
Heating time at 150°C
14 min.
16 min.
18 min.
20 min.
Timer
99:59
hh:min e ∞
99:59 hh:min e ∞
99:59 hh:min e ∞
99:59 hh:min e ∞
Safety class
3.1
3.1
3.1
3.1
Power supply/power
230 V / 700 W
230 V / 1000 W
230 V / 1900 W
230 V / 2100 W
Internal dimensions (L*A*P)
320 x 320 x 285 mm
400 x 420 x 330 mm
520 x 495 x 450 mm
650 x 640 x 495 mm
Shelves number (standard/max)
2/3
2/5
2/6
2/9
Distance between shelves
50 mm
50 mm
50 mm
50 mm
Max loading of shelves
10 Kg
15 Kg
20 Kg
20 Kg
External dimensions (L*A*P)
460 x 685 x 530 mm
690 x 635 x 470 mm
815 x 750 x 600 mm
940 x 905 x 660 mm
Weight
40 Kg
53 Kg
74 Kg
103 Kg
Forced air ovens
TCF-50 Plus
TCF-120 Plus
TCF-200 Plus
TCF-400 Plus
Volume
50 l
120 l
200
400 l
Max temperature/ Resolution
+300/0,1°C
+300/0,1°C
+300/0,1°C
+300/0,1°C
Homogeneity at 150°C
± 2 %
± 2 %
± 2 %
± 2 %
Temperature variation at 150°C
± 0,3°C
± 0,3°C
± 0,4°C
± 0,5°C
Heating time at 150°C
20 min.
24 min.
30 min.
50 min.
Timer
99:59 hh:min e
∞
99:59 hh:min e
∞
99:59 hh:min e
∞
99:59 hh:min e
∞
Safety class
3.1
3.1
3.1
3.1
Power supply/power
230 V / 980 W
230 V / 1900 W
230 V / 2400 W
230 V / 3200 W
Internal dimensions (L*A*P)
400 x 415 x 310 mm
520 x 530 x 435 mm
645 x 650 x 495 mm
1000 x 800 x 500 mm
Shelves number (standard/max)
2/5
2/7
2/9
2/10
Distance between shelves
50 mm
50 mm
50 mm
50 mm
Max loading of shelves
15 Kg
20 Kg
20 Kg
20 Kg
External dimensions (L*A*P)
690 x 635 x 570 mm
810 x 750 x 690 mm
945 x 870 x 755 mm
1285 x 1060 x 750 mm
Weight
54 Kg
74 Kg
103 Kg
160 Kg
COMMAND
DESCRIPTION
The SET/PROG button allows you to set the working parameters and to enter/exit from the
programs.
In combination with the SHIFT key, it allows you to access to menus with password (see
paragraph 10).
The SHIFT button allows you to quickly change the digit (decimal, units, tens, etc.) of the
value of the parameter you are editing.
In combination with the SET/PROG key allows you to access to menus with password (see
paragraph 10).
Adjustment buttons allow you to increase or decrease the value of the operating parameter
being edited.
The START / STOP button allows you to start / stop an operating cycle or a program.
The ON/OFF button allows you to turn on or off the instrument.
Содержание ICF-120 Plus
Страница 1: ...OVENS and INCUBATORS User manual...