UM Ovens and Incubators EN rev. 0 31.05.2021 

 

Delay starts function

 

After selecting the program to which this function is to be applied through the parameter Pn, it is possible to set a 
delay (in hours and minutes) at the start of the operating cycle. 

Set the desired start delay value (hh:mm) by pressing the keys 

You can move quickly between the figures using the SHIFT key 

. Confirm the value by pressing SET/PROG 

again,  the  display  returns  to  the  standby  screen.  By  long  pressing  the  START/STOP  key 

    (4-5  seconds),  the 

instrument starts the program, but does not start heating immediately: the word "end" in the top right-hand corner 
of the display and the delay time flash alternately, marking the wait from the set delay time to the actual start. Once 
the set delay time has elapsed, the instrument starts the program and the regular timer appears on the display. 
 

12.

 

Introduction of samples into oven/incubator 

 

                     DANGER

 

 

 

Explosion hazard and risk of death 

 

Never introduce materials into the instrument that are explosive or flammable at the 

selected operating temperature. 

 

Never  introduce  into  the 

instrument’s

  materials  containing  flammable  or  explosive 

solvents. 

 

Never introduce into the instrument materials that by sublimation or pyrolysis create 

flammable materials at the selected operating temperature. 

 

Poisoning and death hazard 

 

Never introduce materials into the instrument which could create poisonous gases. 

 

Never introduce materials into the instrument that can react with moisture and produce 

explosive gases.

 

 

 

Temperature limit for sample protection

 

The instrument foresees the possibility of limiting the maximum working temperature to protect the samples from 
an incorrect temperature setting of the heating cycle. 

Follow the instructions in paragraph 10 and use the keys 

 to set the password 0003. You can move quickly 

between the digits using the SHIFT key 

. Confirm the value by pressing SET/PROG button 

 again. The display 

in the top right-hand corner shows the parameter "

tm

" (max. temperature) and the maximum value for that type 

of instrument (different between oven and incubator). Set the maximum temperature value that you do not want 

the instrument to exceed during operation by pressing the keys 

. You can move quickly between the digits 

using the SHIFT key 

. Confirm the value by pressing SET/PROG button 

 again. 

NOTE:

 to determine the correct "

tm

" value, the natural and inevitable initial temperature peak that the 

Argolab oven/incubator will have during thermostatting must be considered. 
Application example: If the temperature set for the heating cycle is 100 °C and a limit temperature (tm) of 
70 °C is set, the instrument will attempt to reach the temperature indicated during parameter setting (100 °C), even 
if it is higher than the limit temperature set in this submenu (tm). When 70°C is reached the instrument goes into 
alarm with an intermittent acoustic signal (can be silenced by pressing any key) and the heating element is no longer 
powered until the temperature falls below the limit temperature ("tm"). 

NOTE:

  the  instrument  will  always  attempt  to  reach  the  temperature  set  for  the  heating  cycle  and 

consequently, as long as it is higher than the limit temperature, the device will go into an over-temperature 
alarm as explained in the previous paragraph. 

 
 

Содержание ICF-120 Plus

Страница 1: ...OVENS and INCUBATORS User manual...

Страница 2: ...troller Model Description Temperature range ICN 16 Plus Natural convection incubator 16L maximum volume Room temperature from 5 C to 70 C ICN 35 Plus Natural convection incubator 35L maximum volume Ro...

Страница 3: ...gns 5 Symbols of obligation 5 2 GENERAL SAFETY INSTRUCTIONS 6 3 CE MARKING DATA 8 4 CONTENT OF PACKAGE 9 5 TRANSPORTATION 9 Instructions for safe transportation 9 Transportation of an already used ove...

Страница 4: ...Cycle functions 15 Repetition of a selected program 15 Delay starts function 16 12 INTRODUCTION OF SAMPLES INTO OVEN INCUBATOR 16 Limit temperature for sample protection 16 Restart mode after absence...

Страница 5: ...ions or incorrect results due to failure to comply with them Read entirely and carefully this manual and be sure to familiarize with the tool before starting to work with it This manual must be kept n...

Страница 6: ...surfaces Danger of damage to health caused by toxic substances Danger of injury from overturning of objects Risk of injury from lifting heavy objects Danger of environmental damage Danger of corrosio...

Страница 7: ...y where there is no risk of explosion Do not keep air solvent mixtures or explosive dusts nearby Never introduce materials into the instrument that are explosive or flammable at the selected operating...

Страница 8: ...f each other Always fix 2 stacked ovens with the fixing plates provided Risk of injury Risk of slipping or overturning and risk of damage to the instrument The instrument must be lifted by 2 people Th...

Страница 9: ...ors EN rev 0 31 05 2021 3 CE marking data Argolab instruments are manufactured in compliance with Directive 2006 42 EC and the relevant Community Directives applicable at the time of placing on the ma...

Страница 10: ...rtation ATTENTION Risk of injury of slipping or overturning and risk of damage to the instrument The instrument must be lifted by 2 people The instrument must be transported in its original packaging...

Страница 11: ...rmitted storage temperature is 10 C to 60 C The maximum permitted storage humidity is 85 RH without condensation 7 First installation Getting started The oven should be installed in following conditio...

Страница 12: ...11 UM Ovens and Incubators EN rev 0 31 05 2021 8 Instrument parts Display and commands Display Adjustment keys ON OFF Key Opening closing handle...

Страница 13: ...mperature variation at 150 C 0 3 C 0 3 C 0 4 C 0 5 C Heating time at 150 C 20 min 24 min 30 min 50 min Timer 99 59 hh min e 99 59 hh min e 99 59 hh min e 99 59 hh min e Safety class 3 1 3 1 3 1 3 1 Po...

Страница 14: ...at 37 C 40 min 45 min 55 min Timer 99 59 hh min e 99 59 hh min e 99 59 hh min e Safety class 3 1 3 1 3 1 Power supply power 230 V 600 W 230 V 700 W 230 V 1500 W Internal dimensions L A P 520 x 530 x 4...

Страница 15: ...using the buttons in H High M Medium L Low otherwise go to STEP 2 STEP 2 Use the and SHIFT keys to set the temperature value of the second working STEP Briefly press the SET PROG button to confirm the...

Страница 16: ...e this operation has been carried out the word Lk lock appears in the top right hand corner of the display in place of the word TIME next to the digits 0000 password Here below you can find the passwo...

Страница 17: ...mit for sample protection The instrument foresees the possibility of limiting the maximum working temperature to protect the samples from an incorrect temperature setting of the heating cycle Follow t...

Страница 18: ...ormal and therefore reducing the AL value excessively would risk causing the instrument to go into alarm frequently and unnecessarily Follow the instructions in paragraph 10 and use the keys to set th...

Страница 19: ...corrosive DANGER Risk of electric shock and death DO NOT spray water or cleaning agents onto the interior and exterior surfaces Switch off the main switch and pull out the mains plug before cleaning D...

Страница 20: ...does not apply to any product or parts thereof that have been damaged due to incorrect installation improper connection misuse accident or abnormal operating conditions No liability is accepted for d...

Отзывы: