Program recall
When the instrument is switched on and in standby (heating cycle off), by briefly pressing the SET/PROG key
once, the word "PROG" and the program number alongside start flashing simultaneously.
Use the keys
to call up the desired programme. Confirm by briefly pressing the SET/PROG button
. The
selected program is ready to start.
•
Modify of a program
To change a program, it is necessary to keep pressed the SET/PROG key
for a few seconds: the word "PROG"
and the program number start flashing simultaneously and then only the program number flashes. At this point, it
is possible to choose the program number to be modified using the keys
and confirm the choice by pressing
the SET/PROG key
briefly. Subsequently, the instrument enters edit mode of the program you want to modify
and the temperature value of the first STEP flashes together with the word "PROG", indicating that you are in the
programming phase.
STEP 1
Use the
and SHIFT keys
to set the temperature value of the first working STEP. Briefly press the SET/PROG
button
to confirm the temperature value and to switch to the timer value (STEP 1). Use the keys
and
SHIFT button
to set the desired time value for the first STEP and confirm the value by pressing the SET/PROG
button
shortly. If the instrument has forced ventilation, the next parameter will be the fan speed, which can be
set using the buttons
in (H=High, M=Medium, L=Low), otherwise go to STEP 2.
STEP 2
Use the
and SHIFT keys
to set the temperature value of the second working STEP. Briefly press the
SET/PROG button
to confirm the temperature value and to switch to the timer value (STEP 2). Use the keys
and SHIFT button
to set the desired time value for the STEP 2 and confirm the value by pressing the
SET/PROG button
shortly. If the instrument has forced ventilation, the next parameter will be the fan speed,
which can be set using the buttons
in (H=High, M=Medium, L=Low), otherwise go to STEP 3.
REPEAT THE PREVIOUS INSTRUCTIONS FOR EACH STEP YOU WANT TO PROGRAM.
NOTE:
if you do not want to use all the 10 STEPS of the program you are storing, it is necessary to force
the instrument to finish the program itself. To do this, simply set the time to "00:00" in the step following
the last one you wish to use.
EXAMPLE: If the last working step to be used is the fifth, it is enough to set the timer in the sixth step to
"00:00", thus forcing the instrument to stop at the end of the fifth step.
•
Modify of Prog 0
To modify "PROG 0" select the above-mentioned program in the selection phase and, by keeping pressed the
SET/PROG button
for few seconds, the word PROG and the number 0 will flash simultaneously and then only
the number 0; press SET/PROG button
again to enter the edit mode which allows to set the desired
temperature, timer and fan speed (where applicable).
NOTE:
in "PROG 0" by setting timer 00:00, the Argolab oven/incubator will work at the set temperature
until the operator stops the heating cycle by pressing the START/STOP key
.