WindMaster and WindMaster Pro
1561-PS-0001.
Issue 11 February 2017
________________________________________________________________________________________________________________________
56
Nx – Set unit ID
Sets the unit ID displayed at the start of all GILL communications strings
Set
Q (default),
…from A to Z
Command N<Q>
Ox – Set ASCII output format
1
Comma Separated Variable (CSV) format
2
Fixed field
Example string for CSV data changing to an error status code condition.
Q,335.3,001.261,-001.282,M,+345.41,+023.05,00,+2.4181,+2.4181,+2.4162,+2.4175,-
50.00C, 78
Q,,,,M,,,07,+2.4181,+2.4187,+2.4162,+2.4175,-
50.00C, 72
Example string for Fixed Field data changing to an error status code condition.
Q,251.7,000.860,-000.401,M,+346.43,+024.80,00,+2.4181,+2.4187,+2.4169,+2.4175,-
50.00C, 7F
Q,999.9,999.999,+999.999,M,+999.99,+999.99,07,+2.4181,+2.4187,+2.4169,+2.4181,-
50.00C, 77
Px – Set output rate
Note: the P9 setting is an option on the WindMaster unit, which must be specified at the time of ordering.
1
Rate is 1 Hz (20Hz Sampling)
2
Rate is 2 Hz (20Hz Sampling)
3
Rate is 4 Hz (20Hz Sampling)
4
Rate is 5 Hz (20Hz Sampling)
5
Rate is 8 Hz (32Hz Sampling)
6
Rate is 10 Hz (20Hz Sampling)
7
Rate is 16 Hz (32Hz Sampling)
8
Rate is 20 Hz (20Hz Sampling)
9
Rate is 32 Hz (32Hz Sampling)
20
Rate is ¼ Hz
21 Rate is ½ Hz
Sx – Analogue output range
This sets the velocity scaling for the analogue output. (see also Para 9.5).
1
Full range is 5m/s
2
Full range is 10m/s
3
Full range is 20m/s
4
Full range is 30m/s
5
Full range is 40m/s
6
Full range is 50m/s
7
Full range is 60m/s
8
Full range is 120m/s