WindMaster and WindMaster Pro
1561-PS-0001.
Issue 11 February 2017
________________________________________________________________________________________________________________________
19
6.6.2
WindMaster 1590-PK-050 Pipe Mount Box Connection Details and WindMaster and
WindMaster Pro Pipe Mount Box Connection Details (Spared Item 1590-PK-048).
See also the next page.
Part
WindMaster 1590-PK-050
Pipe Mount Integral Part
Pipe Mount Adaptor 1590-PK-048
Spared Item
Connector
Terminal
Connection Name
Connection Name
J1
PRT (No Connection)
PRT (wired connection)
PRT Excite – (No Connection)
PRT Excite – (wired connection)
PRT Sense + (No Connection)
PRT Sense + (wired connection)
PRT Sense – (No Connection)
PRT Sense – (wired connection)
Screen Ground from WindMaster
Screen Ground from WindMaster
Com (communication) Mode
(External Link to determine RS232 or
RS422/RS485)
Com (communication) Mode
(External Link to determine RS232 or
RS422/RS485)
Power 0V
Power 0V
Power +V
Power +V
J2
Anemometer Transmit TXA (-)
Anemometer Transmit TXA (-)
Anemometer Transmit TXB (+)
Anemometer Transmit TXB (+)
Anemometer Receive RXA (-)
Anemometer Receive RXA (-)
Anemometer Receive RXB (-)
Anemometer Receive RXB (-)
Analogue Out 4 (No Connection)
Analogue Out 4 (wired connection)
Analogue Out 3 (No Connection)
Analogue Out 3 (wired connection)
Analogue Out 2 (No Connection
Analogue Out 2 (wired connection
Analogue Out 1 (No Connection)
Analogue Out 1 (wired connection)
J3
Analogue In 1
Analogue In 1
Analogue Ground
Analogue Ground
Analogue In 2
Analogue In 2
Analogue Ground
Analogue Ground
Analogue In 3
Analogue In 3
Analogue Ground
Analogue Ground
Analogue In 4
Analogue In 4
Analogue Ground
Analogue Ground
J4
(Jumper)
Com-CM Communication mode set by external
connection (Default Setting)
Communication mode set by external
connection (Default Setting)
Com-OV Set communication mode to
RS422/RS485 (Internal Link, external
link unnecessary)
Set communication mode to
RS422/RS485 (Internal Link, external
link unnecessary)
Com-V+ Set communication mode to RS232
(Internal Link, external link
unnecessary)
Set communication mode to RS232
(Internal Link, external link
unnecessary)
Screen
Ground Pillar
Screen
The cable screen from the
WindMaster 31 way connector is
wired to a Ground Pillar
The cable screen from the
WindMaster 31 way connector is
wired to a Ground Pillar
Screen
Ground Pillar
Screen
The cable screen/s from user
equipment is wired to this Pillar using
M4 screw/washer/terminal fixings
supplied in the installation kit
The cable screen/s from user
equipment is wired to this Pillar using
M4 screw/washer/terminal fixings
supplied in the installation kit