WindMaster and WindMaster Pro
1561-PS-0001.
Issue 11 February 2017
________________________________________________________________________________________________________________________
34
Notes:
The M20 Cable gland from the connection box is designed to accept cables with an outer diameter between 8mm to
13mm.
Connect the external cable screen/s to the pillar inside the box using the supplied M4 parts illustrated above.
Appropriate strain relief MUST be fitted to the cable connected through the M20 cable gland - this is particularly
important with pipe mounting.
The weight of the Pipe Mount is approximately 1.12kg.
Grounding (Earthing)
For protection against lightning the WindMaster
MUST
be earthed via its mountings.
The unit MUST be connected to an appropriate grounding point with a minimum of 6mm² copper wire.
An M6 ring terminal is fitted to the WindMaster as illustrated above to which the 6mm² copper wire can be crimped
attached for connection to a ground point.