2.5.3
Commands
The remote commands used in the 1026 syntax are listed in Table 2-8. In the case of a
command requiring a numeric argument, the acceptable range of values is determined by the
specifications of the instrument. For example, it is not permitted to specify a frequency outside
of the range of the instrument.
Command
Description
AF0
fast always off
AF1
fast always on
DISP0
displays off
DISP1
displays on
EXTALCzOFF
disable external automatic level control
EXTALCzON
enable external automatic level control
FAzn
set FA (RF output frequency) to n MHz
HP
select GT 9000 syntax (use GT command to return to the present syntax)
LCO
level correction off
LC1
level correction on
LEVELzn
set output level to n dBm (entry resolution .1 dB)
LVLCRSzn
set attenuator to n dB, independently of leveling loop (entry resolution: 10 dB steps);
output level is the sum of LVLCRS & LVLFNE arguments
LVLFNEzn
set leveling loop to n dB, independently of attenuator (entry resolution:
.1 dB); output level is the sum of LVLCRS & LVLFNE arguments
MODzAM
external amplitude modulation
MODzCONTzOFF
AM/scan modulation off
MODzCONTzSCAN
scan modulation on
MODzCONTzAM
external AM on
MODzDE
external pulse modulation, rising-edge triggered
MODzDESCRzEXT-
external pulse modulation, falling-edge triggered
MODzDESCRzOFF
pulse modulation off
MODzDESCRzPULSE
internal pulse modulation
MODzDESCRzSQR
internal pulse modulation, 50% duty cycle
external pulse modulation, positive trigger
MODzEXT-
external pulse modulation, negative trigger
MODzOFF
modulation off
MODzPULSE
internal pulse modulation, 1 kHz rate, 1
µ
s width
MODzSQR
internal square wave modulation, 1kHz rate
NSR
never say remote (activates front panel controls and displays; instrument accepts
commands via front panel controls AND remote control bus)
PULSEzEXT
external pulse modulation, positive trigger
PULSEzOFF
pulse modulation off
SENDzFREQ
send data to controller, specifying current output frequency or measured frequency
SENDzPOWER
send data to controller, specifying current output power or measured power
SENDzSTATUS
send data to controller, specifying current lock and level status
UNz1
unlocked mode on
Table 2-8. 1026 Syntax Commands
Model GT 9000 Microwave Signal Generator
2-26
Manual No. 120AM00250, Rev C, September 1998
Содержание GT 9000
Страница 12: ...Series GT 9000 Microwave Synthesizers x Manual No 120AM00250 Rev C September 1998 ...
Страница 14: ...Series GT 9000 Microwave Synthesizers xii Manual No 120AM00250 Rev C September 1998 ...
Страница 18: ...Series GT 9000 Microwave Synthesizers xvi Manual No 120AM00250 Rev C September 1998 ...
Страница 20: ...Series GT 9000 Microwave Synthesizers xviii Manual No 120AM00250 Rev C September 1998 ...
Страница 94: ...Model GT 9000 Microwave Synthesizer 3 36 Manual No 120AM00250 Rev C September 1998 ...
Страница 122: ...Model GT 9000 Microwave Synthesizer 4 28 Manual No 120AM00250 Rev C September 1998 ...
Страница 129: ...6 Check the level control circuit see Section 5 3 7 Maintenance Manual No 120AM00250 Rev C September 1998 5 3 ...
Страница 142: ...Model GT 9000 Microwave Synthesizer 5 16 Manual No 120AM00250 Rev C September 1998 ...
Страница 198: ...Model GT 9000 Microwave Synthesizer 6 56 Manual No 120AM00250 Rev C September 1998 ...
Страница 206: ...Model GT 9000 Microwave Synthesizers 7 2 Manual No 120AM00250 Rev C September 1998 ...
Страница 232: ...Model GT 9000 Microwave Synthesizer A 26 Manual No 120AM00250 Rev C September 1998 ...
Страница 238: ...Model GT 9000 Microwave Synthesizer Index 6 Manual No 120AM00250 Rev C September 1998 ...