Procedure
1.
Connect the power sensor to the UUT. Disable the leveling system of the UUT. Step the
frequency across the particular instrument operating range in 50 MHz increments. At
each point measure the power, apply the appropriate sensor correction factor, and plot
the result.
2.
Using the above setup, set the UUT to 0 dBm with the leveling enabled. Again, step the
frequency across the operating range; measure, correct, and plot the power.
3.
For instruments with Option 26:
Connect the UUT to the measuring receiver via the downconverter. With the UUT at
0 dBm and set to the first test frequency, establish a reference on the receiver. Reduce
the UUT output in 10 dB steps, observing and recording the receiver reading. As
needed, perform the recalibration requested by the receiver.
4.
Repeat for the remaining test frequencies.
Model GT 9000 Microwave Synthesizer
4-18
Manual No. 120AM00250, Rev C, September 1998
Содержание GT 9000
Страница 12: ...Series GT 9000 Microwave Synthesizers x Manual No 120AM00250 Rev C September 1998 ...
Страница 14: ...Series GT 9000 Microwave Synthesizers xii Manual No 120AM00250 Rev C September 1998 ...
Страница 18: ...Series GT 9000 Microwave Synthesizers xvi Manual No 120AM00250 Rev C September 1998 ...
Страница 20: ...Series GT 9000 Microwave Synthesizers xviii Manual No 120AM00250 Rev C September 1998 ...
Страница 94: ...Model GT 9000 Microwave Synthesizer 3 36 Manual No 120AM00250 Rev C September 1998 ...
Страница 122: ...Model GT 9000 Microwave Synthesizer 4 28 Manual No 120AM00250 Rev C September 1998 ...
Страница 129: ...6 Check the level control circuit see Section 5 3 7 Maintenance Manual No 120AM00250 Rev C September 1998 5 3 ...
Страница 142: ...Model GT 9000 Microwave Synthesizer 5 16 Manual No 120AM00250 Rev C September 1998 ...
Страница 198: ...Model GT 9000 Microwave Synthesizer 6 56 Manual No 120AM00250 Rev C September 1998 ...
Страница 206: ...Model GT 9000 Microwave Synthesizers 7 2 Manual No 120AM00250 Rev C September 1998 ...
Страница 232: ...Model GT 9000 Microwave Synthesizer A 26 Manual No 120AM00250 Rev C September 1998 ...
Страница 238: ...Model GT 9000 Microwave Synthesizer Index 6 Manual No 120AM00250 Rev C September 1998 ...