ALC (6)
The use of the ALC (automatic level control) function allows the ALC circuit to take a level
feedback from an external detector or power meter DC-output (see the rear panel External ALC
connector).
Special (7)
The SPECIAL feature provides access to numbered special functions. When you press [SHIFT]
[SPECIAL], the Entry Menu displays
ENTER SPECIAL #
or
SCROLL
. Use the [SCROLL] key to
review the available functions listed below:
00
Erase Memories. Press [ON] to erase stored setup memories 0 to 9.
01
Model & Version. Press [ON] to display the model number, frequency range, and
software version.
02
Display Bright. Press [ON] to adjust the brightness of the entry display.
11
Ext Ref. Select 5 MHz or 10 MHz external reference (see Option 11).
20
RF Off @ PWR Up. This function controls RF output power upon power-up. Press
[ON] or [OFF] to enable or disable this function.
21
RF ON @ FREQ CNG. This function controls RF power output during frequency
changes, except during band or filter changes. Press [ON] or [OFF] to enable or
disable this function.
40
HP GPIB Syntax. Press [ON] to use the normal (GT 9000) remote control command
set.
41
GT GPIB Syntax. Press [ON] to use the Giga-tronics 1026 remote control command
set.
50
Yig Fast Mode. Press [ON] or [OFF] to enable or disable (this function is used only
for test purposes).
51
Unlocked Mode. Press [ON] or [OFF] to enable or disable (this function is used only
for test purposes).
99
Clear Specials. Press [ON] to reset all special functions to the default mode.
Recall/Store
Up to ten front panel setups can be stored in non-volatile memory. After the desired parameters
have been set (or allowed to default), the setting can be stored by first pressing [SHIFT]
followed by [STORE] and then the desired memory number (0 to 9). The data is stored
immediately. To recall a setup, press [RECALL] (without pressing the SHIFT key) followed by
the memory number (0 to 9).
Operation
Manual No. 120AM00250, Rev C, September 1998
2-7
Содержание GT 9000
Страница 12: ...Series GT 9000 Microwave Synthesizers x Manual No 120AM00250 Rev C September 1998 ...
Страница 14: ...Series GT 9000 Microwave Synthesizers xii Manual No 120AM00250 Rev C September 1998 ...
Страница 18: ...Series GT 9000 Microwave Synthesizers xvi Manual No 120AM00250 Rev C September 1998 ...
Страница 20: ...Series GT 9000 Microwave Synthesizers xviii Manual No 120AM00250 Rev C September 1998 ...
Страница 94: ...Model GT 9000 Microwave Synthesizer 3 36 Manual No 120AM00250 Rev C September 1998 ...
Страница 122: ...Model GT 9000 Microwave Synthesizer 4 28 Manual No 120AM00250 Rev C September 1998 ...
Страница 129: ...6 Check the level control circuit see Section 5 3 7 Maintenance Manual No 120AM00250 Rev C September 1998 5 3 ...
Страница 142: ...Model GT 9000 Microwave Synthesizer 5 16 Manual No 120AM00250 Rev C September 1998 ...
Страница 198: ...Model GT 9000 Microwave Synthesizer 6 56 Manual No 120AM00250 Rev C September 1998 ...
Страница 206: ...Model GT 9000 Microwave Synthesizers 7 2 Manual No 120AM00250 Rev C September 1998 ...
Страница 232: ...Model GT 9000 Microwave Synthesizer A 26 Manual No 120AM00250 Rev C September 1998 ...
Страница 238: ...Model GT 9000 Microwave Synthesizer Index 6 Manual No 120AM00250 Rev C September 1998 ...