Remote Operation
Manual 31470, Rev. E, April 2001
3-105
3.19.6
Measured Offset Entry
A measurement can be saved and used as an offset. The command format for this purpose is:
Syntax
:
[AP, BP, AR, BR, AD, or BD] OS DO EN
The command begins with a function code which describes the measurement that is to be
stored as an offset value. There are six possible function codes; they are interpreted as
follows:
AP:A
BP:B
AR:A/B
BR:B/A
AD:A-B
BD:B-A
OS followed by DO indicates that the difference between the current offset and the current
value of the measurement described in the prefix, is to be saved as an offset value.
A terminating suffix (EN) is required.
Examples:
OUTPUT 713;AP OS DO EN
! Save measurement A as an offset
OUTPUT 713;BP OS DO EN
! Save measurement B as an offset
OUTPUT 713;AR OS DO EN
! Save measurement A/B as an offset
OUTPUT 713;BR OS DO EN
! Save measurement B/A as an offset
OUTPUT 713;AD OS DO EN
! Save measurement A-B as an offset
OUTPUT 713;BD OS DO EN
! Save measurement B-A as an offset
Содержание 8650A Series
Страница 4: ......
Страница 12: ...Series 8650A Universal Power Meters viii Manual 31470 Rev E April 2001 ...
Страница 16: ...Series 8650A Universal Power Meters xii Manual 31470 Rev E April 2001 ...
Страница 18: ...Series 8650A Universal Power Meters xiv Manual 31470 Rev E April 2001 ...
Страница 20: ...Series 8650A Universal Power Meters xvi Manual 31470 Rev E April 2001 ...
Страница 32: ...Series 8650A Universal Power Meters 1 12 Manual 31470 Rev E April 2001 ...
Страница 198: ...Series 8650A Universal Power Meters 3 124 Manual 31470 Rev E April 2001 ...
Страница 216: ...Series 8650A Universal Power Meters A 8 Manual 31470 Rev E April 2001 ...
Страница 228: ...Series 8650A Universal Power Meters B 12 Manual 31470 Rev E April 2001 ...
Страница 240: ...Series 8650A Universal Power Meters Index 6 Manual 31470 Rev E April 2001 ...