Series 8650A Universal Power Meters
3-42
Manual 31470, Rev. E, April 2001
3.2.23
Limit Line Configuration and Monitoring
CALCulate:LIMit
The CALCulate:LIMit commands specify and query the status of power measurement limit values and
limit line pass/fail checking. This allows the 8650A to monitor measured values and determine if the
values are outside certain limits or above/below a single limit. The upper limit cannot be specified any
lower than the lower limit; meaning that an exclusion zone of values cannot be specified.
Limit values can be specified separately for either of the two software calculation channels - 1 and 2.
These channels can be specified to correspond to power sensors 1 and 2 (default) or as a combination of
the power sensors. For example, set limit lines to monitor overload or under range conditions on an
amplifier’s output power and gain by specifying channel 1 as sensor 1 limits checking and channel 2 as
Sensor 2/Sensor 1. Power sensor assignments to channel measurement definition is also part of the
CALCulate Subsystem. Limit monitoring can not be performed for individual sensors unless a channel
is configured for single sensor, POW 1 or POW 2 operation.
Upper and Lower limits can be active simultaneously.
Unless serial requests are enabled, the CALCulate:LIMit commands can not be configured to
automatically alert the controller directly during a limit violation; the 8650A must be queried to receive
information regarding pass/fail status of limit line violations.
Automatic notification of limit line violations is accomplished using the status byte and the operation
register. A controller can be notified of a limit line violation via the request service, *SRE command.
After the controller receives the request service, query the event status register. Check the status register
for a limit line violation or send CALCulate:LIMit:FAIL? to check if a limit line is being or has been
violated. 1 indicates a limit line violation. 0 indicates the channel measurement is within the limit
lines.
Syntax
:
CALCulate<channel 1 or 2>:LIMit:CLEar[:IMMediate]
CALCulate - LIMit
Example
:
OUTPUT @Pwr_mtr;CALC1:LIM:CLE
! Reset channel 1 limit violation indicator to 0
Description
: This command resets the limit line pass/fail indicator to 0.
Syntax
:
CALCulate<channel 1 or 2>:LIMit:FCOunt?
CALCulate - LIMit FCOunt?
Example
:
OUTPUT @Pwr_mtr;CALC1:LIM:FCO?
! Query number of channel 1 limit failures
Response
:
Single number; the number of times the limit lines were exceeded.
Description
: This command reports the number of times a limit line has been exceeded since the limit
checking function was set to ON, or the limit line monitor was cleared with
CALC#:LIM:CLE.
Содержание 8650A Series
Страница 4: ......
Страница 12: ...Series 8650A Universal Power Meters viii Manual 31470 Rev E April 2001 ...
Страница 16: ...Series 8650A Universal Power Meters xii Manual 31470 Rev E April 2001 ...
Страница 18: ...Series 8650A Universal Power Meters xiv Manual 31470 Rev E April 2001 ...
Страница 20: ...Series 8650A Universal Power Meters xvi Manual 31470 Rev E April 2001 ...
Страница 32: ...Series 8650A Universal Power Meters 1 12 Manual 31470 Rev E April 2001 ...
Страница 198: ...Series 8650A Universal Power Meters 3 124 Manual 31470 Rev E April 2001 ...
Страница 216: ...Series 8650A Universal Power Meters A 8 Manual 31470 Rev E April 2001 ...
Страница 228: ...Series 8650A Universal Power Meters B 12 Manual 31470 Rev E April 2001 ...
Страница 240: ...Series 8650A Universal Power Meters Index 6 Manual 31470 Rev E April 2001 ...