Series 8650A Universal Power Meters
3-32
Manual 31470, Rev. E, April 2001
Syntax
:
SENSe<sensor 1 or 2>:TRIGger:LEVel[:MAGnitude]
space
<R In EXT triggering use -0.1
to 5.000 V. In INT triggering use -30 to +20 dBm>
SENSe:TRIGger - LEV
Example
:
OUTPUT @Pwr_mtr;SENS1:TRIG:LEV 1.7! Set peak sensor 1 trigger level to1.7 volts
Description
: This command sets the trigger level for the 80350A Series Peak Power Sensors. Under
INTernal triggering the value is in dBm (-30 to +20 dBm, with default of -20 dBm). When
EXTernal triggering is used the value is in Volts (-0.100 to 5.000 V, with a default of 1.700
V). The trigger level must be lower than the signal that is triggering the peak power sensor.
Under Internal triggering, this means that a -16 dBm signal must be triggered with a
setting that is less than -16 dBm. EXT responds similarly. The input impedance to the EXT
trigger port on the back of the sensor housing is not 50 ohms: it is a high impedance to
prevent damage from triggering inputs such as 5.0 V TTL signals. If you use a high speed
50
Ω
trigger source, this may cause unwanted signal reflections and noise on the trigger
line. An SMB to SMB attenuator is available from Giga-tronics to alleviate this condition.
The attenuation reduces the reflected noise, and reduces/eliminates the noisy trigger
characteristics. (80340 Series Peak Power Sensors do not support this command.)
Be sure to set the trigger level after the INTernal or EXTernal triggering is selected.
Syntax
:
SENSe<sensor 1 or 2>:TRIGger:OFFSet[:MAGnitude]
SENSe:TRIGger - OFFSET
Example
:
OUTPUT @Pwr_mtr;SENS1:TRIG:OFFS 0
! Set peak sensor 1 trigger offset time to 0
! second
Description
: This command sets the peak power sensor trigger offset time.
(The 80340 Series Power Sensors do not support this command.)
Syntax
:
SENSe<sensor 1 or 2>:TRIGger:TOTAl[:MAGnitude]?
SENSe:TRIGger - TOTAI
Example
:
OUTPUT @Pwr_mtr;SENS1:TRIG:TOTA?
! Query peak sensor 1 total trigger delay
! time
Response
:
Current peak power sensor total trigger delay setting; a single value in seconds
Description
: This query reports the peak power sensor total trigger delay time
Содержание 8650A Series
Страница 4: ......
Страница 12: ...Series 8650A Universal Power Meters viii Manual 31470 Rev E April 2001 ...
Страница 16: ...Series 8650A Universal Power Meters xii Manual 31470 Rev E April 2001 ...
Страница 18: ...Series 8650A Universal Power Meters xiv Manual 31470 Rev E April 2001 ...
Страница 20: ...Series 8650A Universal Power Meters xvi Manual 31470 Rev E April 2001 ...
Страница 32: ...Series 8650A Universal Power Meters 1 12 Manual 31470 Rev E April 2001 ...
Страница 198: ...Series 8650A Universal Power Meters 3 124 Manual 31470 Rev E April 2001 ...
Страница 216: ...Series 8650A Universal Power Meters A 8 Manual 31470 Rev E April 2001 ...
Страница 228: ...Series 8650A Universal Power Meters B 12 Manual 31470 Rev E April 2001 ...
Страница 240: ...Series 8650A Universal Power Meters Index 6 Manual 31470 Rev E April 2001 ...