Series 8650A Universal Power Meters
3-94
Manual 31470, Rev. E, April 2001
The time value (numeric variable [t]) is an optional variable which specifies a fixed delay
between measurements. The time value specifies the time (in ms) to wait between
measurements; the minimum value is zero. The maximum value is 5000 ms (five
seconds).
Examples
:
OUTPUT 713;FBUF PRE GET BUFFER 200
! take measurements (as fast as possible)
! until GET is received
! then output the last 200 measurements
! taken
OUTPUT 713;FBUF POST TTL BUFFER 100 TIME 2
! wait for a TTL trigger, then take readings
! at intervals of 2 ms
! until a total of 100 measurements have
! been taken
Two simpler commands are also based on the FBUF function code:
Syntax
:
FBUF [DUMP or OFF]
[DUMP] stops the data measurement and buffering, and prepares to return the data taken
so far to the host, even if fewer than the requested number of measurements have been
taken. The requested number of measurements are still returned to the host (the extra
measurements beyond those actually taken are represented by the number -300.00).
[OFF] causes the 8650A to exit the fast buffered mode. All unread data is lost.
Examples
:
OUTPUT 713;FBUF DUMP
! Stop measurement and buffering
OUTPUT 713;FBUF OFF
! Exit the fast buffered mode
Notes on Speed in the Fast Buffered Mode
The fast buffered mode is the fastest method of collecting measurement data. Top speed in the fast
buffered mode is achieved by using a low averaging number (
≤
4), the POST trigger mode, and no time
delay between measurements.
The POST trigger mode is faster than the PRE trigger mode because in the latter mode the 8650A must
check for a trigger between each measurement. In the POST mode, the 8650A is in freerun operation
after the trigger is received.
After issuing the Fast Buffered command, allow for the 8650A to set up in Fast Buffered mode before
issuing a TTL or GET (Group Executable Trigger). Typical delay may vary from 200 msec to 500 msec
based on the number of samples to be collected.
NOTE:
This delay is in addition to the relatively short time it takes to
perform each measurement. If no time value is specified, [t] is assumed to
be zero, and the measurements are taken as fast as possible.
☛
☛
☛
☛
Содержание 8650A Series
Страница 4: ......
Страница 12: ...Series 8650A Universal Power Meters viii Manual 31470 Rev E April 2001 ...
Страница 16: ...Series 8650A Universal Power Meters xii Manual 31470 Rev E April 2001 ...
Страница 18: ...Series 8650A Universal Power Meters xiv Manual 31470 Rev E April 2001 ...
Страница 20: ...Series 8650A Universal Power Meters xvi Manual 31470 Rev E April 2001 ...
Страница 32: ...Series 8650A Universal Power Meters 1 12 Manual 31470 Rev E April 2001 ...
Страница 198: ...Series 8650A Universal Power Meters 3 124 Manual 31470 Rev E April 2001 ...
Страница 216: ...Series 8650A Universal Power Meters A 8 Manual 31470 Rev E April 2001 ...
Страница 228: ...Series 8650A Universal Power Meters B 12 Manual 31470 Rev E April 2001 ...
Страница 240: ...Series 8650A Universal Power Meters Index 6 Manual 31470 Rev E April 2001 ...