5.3.1
Calibration Equipment Required
Test equipment required to calibrate the Models 8501A and 8502A power meters is listed in Table 5-2.
Equivalent or similar test equipment can be substituted.
Description
Representative Model
Key Characteristics
Digital Voltmeter (DVM)
Fluke Model 8800
5 digit display, 0.01% accuracy
Frequency Counter
EIP Model 548
10 digit readout. At least 1 GHz
Frequency Counter/Timer
Tektronix Model DC509
30
µ
s period, 1 ppm
Oscilloscope (Scope)
Tektronix Model 465
5 mV/DIV & 50 MHz bandwidth
Power Meter
HP Model 432
0.95 to 1.05 GHz at 0 dBm, 1.3%
Precision Voltage Source
Digitec Model 3110
5.000V with 0.02% accuracy
Pulsed RF Signal Generator
Giga-tronics Model 907
Trig. Modulator
Variac
General Radio Model W5MT3A
Metered
RF Detector
Giga-tronics Model 16936
High Speed (15 ns)
5.3.2
Preset Conditions
Before stating the calibration procedures, the PPM should be placed into a preset condition. The
Information given in this chapter can be used to calibrate and/or confirm calibration of the PPM and its
detector(s) using the built-in automatic detector calibration system.
A. Frequency Counter/Timer
1.
Set FUNCTION switch to PERIOD A.
2.
Set AVERAGES switch to 10 to the 6th power.
3.
Set SLOPE switch to +.
4.
Set ATTEN switch to X1.
5.
Set SOURCE switch to EXT.
6.
Set COUPL switch to dc.
7.
Set TRIG LEVEL switch to AUTO.
B. Signal Generator
1.
Select MODE - PM (Pulse Modulation).
2.
Select PULSE MODE - INT -PULSE.
3.
Select RATE (HZ) - 1K.
4.
Select PULSE DELAY - X1.
5.
Select PULSE WIDTH - X1.
6.
Adjust RATE to midway between MIN and MAX.
7.
Adjust DELAY to approx. 3
µ
s (minimum).
8.
Adjust WIDTH to a point midway between 0.2
µ
s and 10
µ
s (approx. 5
µ
s).
9.
Connect a cable from the SYNC NORMAL OUTPUT of the Signal Generator to the TRIG
INPUT on the back of the PPM.
10.
Connect the PPM to the Variac, set the Variac to 115 Vac, and turn the power on.
Table 5-2. Required Calibration Test Equipment
Series 8500A Peak Power Meters
5-10
Manual No. 20790, Rev C, November 1998
Содержание 8501A
Страница 12: ...Series 8500A Peak Power Meters x Manual No 20790 Rev C November 1998 ...
Страница 16: ...Series 8500A Peak Power Meters xiv Manual No 20790 Rev C November 1998 ...
Страница 18: ...Series 8500A Peak Power Meters xvi Manual No 20790 Rev C November 1998 ...
Страница 74: ...Series 8500A Peak Power Meters 2 48 Manual No 20790 Rev C November 1998 ...
Страница 156: ...Series 8500A Peak Power Meters 4 30 Manual No 20790 Rev C November 1998 ...
Страница 165: ...Figure 5 3 Model 8501A 8502A Calibration Components Testing and Calibration Manual No 20790 Rev C November 1998 5 9 ...
Страница 180: ...Series 8500A Peak Power Meters 5 24 Manual No 20790 Rev C November 1998 ...
Страница 230: ...Series 8500A Peak Power Meters 8 2 Manual No 20790 Rev C November 1998 ...
Страница 236: ...Series 8500A Peak Power Meters A 6 Manual No 20790 Rev C November 1998 ...
Страница 247: ...Figure D 2 Pre Amplifier Line Driver Schematic RF Detectors Manual No 20790 Rev C November 1998 D 5 ...