5.3.6
A6/A7 Analog Board
Analog PC Board Nos. A6 and A7 are identical. The Model 8501A uses just the A6 board; the Model
8502A uses both A6 and A7. When testing a Model 8502A, all A6 reference designations listed in
this section will become A7 when the A7 board is tested.
A. +10 V Reference Adjust
1.
Connect the DVM LO lead to A6TP1 (REF COMMON).
2.
Connect the DVM HI lead to A6TP2 (+10 V).
3.
Adjust A6R13 (10V REF ADJ) for 10.000V
±
3 mV.
4.
If the unit is an 8502A, repeat Steps 1 through 3 for the A7 Analog board.
5.
Disconnect the DVM.
B. Sampler Zero Adjust.
1.
Set the Peak Averaging Number to 20 by pressing [MENU] (2) [F2] [20] [UNITS]
2.
Connect the sync output from the back of the PPM to the Ext Trig IN on the scope, the Scope HI
lead to A6TP5, and the scope ground lead to A6TP11.
3.
Set the pulse generator to generate a 10
µ
s wide pulse with repetition rate of 1 kHz and
amplitude at 5 V (TTL). Connect the pulse generator to the TRIG INPUT (TTL) on the rear
panel of the PPM.
4.
Press [PEAK] [MENU] [F2].
5.
Adjust A6R112 (SAMPLER ZERO) for 0V
±
2 mV offset at the end of the sample time.
C. Trigger Offset Adjust.
1.
Set the trigger to channel A, Internal, -15 dBm by pressing [MENU] (1) [F1] (channel A if unit
is an 8502A) [-15] [dBm].
2.
Connect the DVM HI lead to A6TP13 (TRIGGER OFFSET).
3.
Adjust A6R113 (TRIG OUT DC OFFSET) for 0V
±
1 mV offset.
4.
If the unit is an 8502A, repeat Steps B and C for board A7.
5.
Disconnect the DVM test leads.
D. Analog Output Adjust.
1.
Press the RESET button on the back of the PPM.
2.
Press [MENU] (11) [F1] [1] [F2].
☛
NOTE: If the PPM required detector calibration when the PPM prompts CONNECT
DETECTOR TO CALIBRATOR, THEN PRESS ANY UNITS KEY, press [7] to access the
service functions.
3.
Connect a cable from the DVM to the ANALOG OUTPUT on the back of the PPM (ANALOG A
output if the unit is an 8502).
4.
Adjust A6R47 (LIN GAIN) for 0.000V
±
10 mV reading on the DVM.
5.
Press [F3].
6.
The DVM should read 3.000V
±
25 mV.
Testing and Calibration
Manual No. 20790, Rev C, November 1998
5-13
Содержание 8501A
Страница 12: ...Series 8500A Peak Power Meters x Manual No 20790 Rev C November 1998 ...
Страница 16: ...Series 8500A Peak Power Meters xiv Manual No 20790 Rev C November 1998 ...
Страница 18: ...Series 8500A Peak Power Meters xvi Manual No 20790 Rev C November 1998 ...
Страница 74: ...Series 8500A Peak Power Meters 2 48 Manual No 20790 Rev C November 1998 ...
Страница 156: ...Series 8500A Peak Power Meters 4 30 Manual No 20790 Rev C November 1998 ...
Страница 165: ...Figure 5 3 Model 8501A 8502A Calibration Components Testing and Calibration Manual No 20790 Rev C November 1998 5 9 ...
Страница 180: ...Series 8500A Peak Power Meters 5 24 Manual No 20790 Rev C November 1998 ...
Страница 230: ...Series 8500A Peak Power Meters 8 2 Manual No 20790 Rev C November 1998 ...
Страница 236: ...Series 8500A Peak Power Meters A 6 Manual No 20790 Rev C November 1998 ...
Страница 247: ...Figure D 2 Pre Amplifier Line Driver Schematic RF Detectors Manual No 20790 Rev C November 1998 D 5 ...
Страница 262: ...Series 8500A Peak Power Meters Index 8 Manual No 20790 Rev C November 1998 ...