Command
Argument
MINPsnn.nn
Set Power for Underrange Condition / none
(s = polarity; n = power in dBm; -15.00 max -50.00 min)
NOTE: Some of the following marker commands are no longer necessary when
writing code pertaining to markers for the 8500A instruments. See Section 3.3.7
MKCAn
(No Op command with the 8500A)
Clear Marker Number Selected on Channel A / none
(n = marker number to clear; 1 to 4)
MKCBn
(No Op with the 8500A)
Clear Marker Number Selected on Channel B / none
(n = marker number to clear; 1 to 4)
MKDAn,dddddd.dd
(No Op with the 8500A) Select Channel A Marker Delay number n and set the delay
to d / none
(n = marker number to clear; 1 to 4; d = marker delay in microseconds; 200,000
max)
MKDBn,dddddd.dd
(No Op with the 8500A) Select Channel B Marker Delay number n and set the delay
to d / none (See MKDA)
MKDFa,b
Place Markers (a and b) which were previously selected and Output Their Delay
Difference (a - b) Marker differences can also be de-selected - See MKDF-1 / When
the PPM is in channel A it will output:
MDFA,a-b,nn.nnnEsnn
(a = first marker; b = second marker; result will be the marker a delay minus the
marker b delay)
MKDF-1
(No Op with the 8500A) Deselect all Marker Difference Pairs / none
MKPA
Autoplace All Four Markers as previously selected and Output Their Delays / If the
PPM is in Channel A it will output:
MRKA1,nn.nnnEsnn;2,nn.nnnEsnn;3,nn.nnnEsnn;4,nn.nnnEsnn
If the PPM is in channel B it will output the same as above only with MRKB header.
(1, 2, 3 and 4 = marker numbers; n = delay in seconds; Esnn = exponent)
MKPFa,xx.x
Place Given Marker on Falling Slope of Pulse Profile at the Selected % Power
Reference and Output Its Delay / When the PPM is in channel A it will output:
MRKAa,nn.nnnEsnn
When the PPM is in channel B it will output:
MRKBa,nn.nnnEsnn
(a = marker number; 1 to 4; x = percent of reference power to place marker; 0.1%
min 99.9% max; n = delay in seconds; Esnn = exponent)
MKPRa,xx.x
Place Given Marker on Rising Slope of Pulse Profile at the Selected % Power
Reference and Output Its Delay / When in channel A the PPM will output:
MRKAa,nn.nnnEsnn
When in channel B the PPM will output:
MRKBa,nn.nnnEsnn
(See MKPF)
MRKA
Select Channel A Marker Sub-Mode / none
MRKB
Select Channel B Marker Sub-Mode / none
Series 8500A Peak Power Meters
3-46
Manual No. 20790, Rev C, November 1998
Содержание 8501A
Страница 12: ...Series 8500A Peak Power Meters x Manual No 20790 Rev C November 1998 ...
Страница 16: ...Series 8500A Peak Power Meters xiv Manual No 20790 Rev C November 1998 ...
Страница 18: ...Series 8500A Peak Power Meters xvi Manual No 20790 Rev C November 1998 ...
Страница 74: ...Series 8500A Peak Power Meters 2 48 Manual No 20790 Rev C November 1998 ...
Страница 156: ...Series 8500A Peak Power Meters 4 30 Manual No 20790 Rev C November 1998 ...
Страница 165: ...Figure 5 3 Model 8501A 8502A Calibration Components Testing and Calibration Manual No 20790 Rev C November 1998 5 9 ...
Страница 180: ...Series 8500A Peak Power Meters 5 24 Manual No 20790 Rev C November 1998 ...
Страница 230: ...Series 8500A Peak Power Meters 8 2 Manual No 20790 Rev C November 1998 ...
Страница 236: ...Series 8500A Peak Power Meters A 6 Manual No 20790 Rev C November 1998 ...
Страница 247: ...Figure D 2 Pre Amplifier Line Driver Schematic RF Detectors Manual No 20790 Rev C November 1998 D 5 ...